Predictika · Laboratory instrument configuration: rheometer, temperature device and measuring system
An AI sales agent prepares a configuration for a modular compact rheometer: the instrument model, a Peltier temperature device, the measuring system geometry and the measurements the customer needs to make. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model of the published model and device limits and returns a verdict, every reason and the figures it computed.
This validator models configurations of one instrument family, the Anton Paar MCR 303, MCR 503 and MCR 503 Power rheometers with Peltier temperature devices, not just the scenarios shown below.
Any configuration of these three models with a Peltier device is in scope — not only the scenarios pictured here. Not covered: prices (no public price list was found, so this validator produces no price build-up or line-item quote), other temperature devices (electrical, liquid, pressure cells, UV and others), maximum frequency of the MCR 503 Power (not listed), other MCR models, accessories, service and regional terms.
7 scenarios in 3 groups · each group runs from the agent’s first attempt to the outcome · click one to switch
Illustrative agent statement. The validator receives the structured request below.
Proposing agent: AI sales agent (scripted)
Before sending, the agent must call Recorded run
What this means
Our assessment of what a typical rules-based CPQ (product rules, price rules and validation rules, as in Salesforce CPQ) does with the same cases. It is a judgment about effort and reliability, not a claim that nothing can be built.
| What the validator caught | Typical rules-based CPQ | Predictika |
|---|---|---|
| Limits that depend on the model chosen | Only with care Torque, speed and frequency limits differ by model, so each limit needs a per-model rule that must stay in step as models are added. | The model and the measurement needs arrive together; every limit is checked against the chosen model on every call. |
| A temperature range that depends on device and geometry | Only with workarounds The dual-drive device has a different top temperature with plate and with cylinder geometries, so the range depends on two choices at once and needs a cross-line rule. | The device and the geometry are in one request; the range is computed from both and the sample temperatures are checked against it. |
| Measuring system that must suit the device | Only with care Plate systems take plate geometries and cylinder systems take cylinder geometries; a product rule can enforce it, but each new device needs its own rule. | Device and geometry are checked together on every call, whichever order they arrive in. |
| Stated temperature range that must match the device | Only with care A stated figure is a free-entry field; a rule can compare it with the computed range, but someone has to write and maintain that comparison. | The stated range travels with the configuration and is compared with the computed range on every call. |
| Every problem returned in one answer | Only with workarounds Rules tend to fire as the order is saved or a step is completed, so a rep often fixes one error and meets the next. | All findings come back together, whatever order the choices arrived in. |
Every rule is checked on every call, in any order the choices arrive; one problem never hides another. Blocking findings stop the quote; approval findings need a person to sign off; everything else is computed.
| 1 | The lowest torque to be measured is 4 nNm; the selected model has a lower limit of 5 nNm (MCR 303), 1 nNm (MCR 503) or 100 nNm (MCR 503 Power) in rotation. |
| 2 | The lowest torque to be measured is 0 nNm; the selected model has a lower limit of 5 nNm (MCR 303), 1 nNm (MCR 503) or 100 nNm (MCR 503 Power) in rotation. |
| 3 | The lowest torque to be measured is 99 nNm; the selected model has a lower limit of 5 nNm (MCR 303), 1 nNm (MCR 503) or 100 nNm (MCR 503 Power) in rotation. |
| 4 | The highest torque to be measured is 231 mNm; the selected model has an upper limit of 230 mNm (MCR 303), 255 mNm (MCR 503) or 300 mNm (MCR 503 Power). |
| 5 | The highest torque to be measured is 256 mNm; the selected model has an upper limit of 230 mNm (MCR 303), 255 mNm (MCR 503) or 300 mNm (MCR 503 Power). |
| 6 | The highest torque to be measured is 301 mNm; the selected model has an upper limit of 230 mNm (MCR 303), 255 mNm (MCR 503) or 300 mNm (MCR 503 Power). |
| 7 | The angular velocity needed is 315 rad/s; the maximum is 314 rad/s (MCR 303, MCR 503) or 200 rad/s (MCR 503 Power). |
| 8 | The angular velocity needed is 201 rad/s; the maximum is 314 rad/s (MCR 303, MCR 503) or 200 rad/s (MCR 503 Power). |
| 9 | The oscillation frequency needed is 101 Hz; the MCR 303 maximum is 100 Hz. |
| 10 | The oscillation frequency needed is 251 Hz; the MCR 503 maximum is 250 Hz. |
| 11 | The Peltier temperature device and the measuring system geometry do not belong together: plate systems take cone-plate or parallel-plate geometries, cylinder systems take concentric-cylinder or double-gap geometries. |
| 12 | The highest sample temperature is 151 degrees C; the selected temperature device reaches 150 degrees C. |
| 13 | The lowest sample temperature is -51 degrees C; the selected temperature device starts at -50 degrees C. |
| 14 | The highest sample temperature is 221 degrees C; the selected temperature device reaches 220 degrees C. |
| 15 | The lowest sample temperature is -11 degrees C; the selected temperature device starts at -10 degrees C. |
| 16 | The highest sample temperature is 51 degrees C; the selected temperature device reaches 50 degrees C. |
| 17 | The quote states a device range of -50 to 200 degrees C; the selected temperature device gives -50 to 220 degrees C. |
| 18 | The quote states a device range of -30 to 220 degrees C; the selected temperature device gives -50 to 220 degrees C. |
| 19 | TruStrain and TruChirp are optional on the MCR 303 (included on the MCR 503 and MCR 503 Power); a person should confirm they are added to the order. |
| 20 | The highest torque to be measured is 300 mNm; the selected model has an upper limit of 230 mNm (MCR 303), 255 mNm (MCR 503) or 300 mNm (MCR 503 Power). |
| 21 | The oscillation frequency needed is 200 Hz; the MCR 303 maximum is 100 Hz. |
| 22 | The highest sample temperature is 250 degrees C; the selected temperature device reaches 220 degrees C. |
The recorded result of the validator for each scenario on this page: the request the AI agent sent, the verdict, and every finding, in the order they appear above.
The agent's request. Please quote an Anton Paar MCR 303 rheometer with an air-cooled plate Peltier system and a concentric-cylinder measuring system, for measurements down to 2 nNm and up to 100 mNm, at up to 100 rad/s and 150 Hz, with samples between -30 and 150 degrees C. The quote states a temperature device range of -10 to 220 degrees C. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote an Anton Paar MCR 303 rheometer with a fluid-cooled plate Peltier system and a cone-plate measuring system, for measurements down to 20 nNm and up to 100 mNm, at up to 100 rad/s and 80 Hz, with samples between -30 and 150 degrees C. The quote states a temperature device range of -50 to 220 degrees C. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote an Anton Paar MCR 503 rheometer with a dual-drive Peltier system and a double-gap measuring system, for measurements down to 500 nNm and up to 100 mNm, at up to 100 rad/s and 200 Hz, with samples between -20 and 170 degrees C. The quote states a temperature device range of -20 to 150 degrees C. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote an Anton Paar MCR 503 rheometer with a dual-drive Peltier system and a double-gap measuring system, for measurements down to 500 nNm and up to 100 mNm, at up to 100 rad/s and 200 Hz, with samples between -20 and 150 degrees C. The quote states a temperature device range of -20 to 150 degrees C. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote an Anton Paar MCR 303 rheometer with a fluid-cooled cylinder Peltier system and a concentric-cylinder measuring system, for measurements down to 20 nNm and up to 100 mNm, at up to 100 rad/s and 80 Hz, with samples between -30 and 200 degrees C. The customer wants TruStrain and TruChirp. The quote states a temperature device range of -50 to 220 degrees C. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote an Anton Paar MCR 303 rheometer with a fluid-cooled cylinder Peltier system and a concentric-cylinder measuring system, for measurements down to 20 nNm and up to 100 mNm, at up to 100 rad/s and 80 Hz, with samples between -30 and 200 degrees C. The customer wants TruStrain and TruChirp. The quote states a temperature device range of -30 to 200 degrees C. It is ready to send.
Verdict: Needs approval.
The agent's request. Please quote an Anton Paar MCR 503 rheometer with a fluid-cooled cylinder Peltier system and a concentric-cylinder measuring system, for measurements down to 20 nNm and up to 100 mNm, at up to 100 rad/s and 80 Hz, with samples between -30 and 200 degrees C. The customer wants TruStrain and TruChirp. The quote states a temperature device range of -30 to 200 degrees C. It is ready to send.
Verdict: Allowed.
The text of the narrated walkthrough, step by step.