Predictika · Laboratory automation configuration: automated cryogenic sample store, labware and installation site
An AI sales agent prepares a configuration for an automated sample store: the BioStore model, the labware to be stored, the temperature setpoint and the room it will be installed in. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model of the published store specifications and returns a verdict, every reason and the figures it computed.
This validator models configurations of one product line, the Azenta BioStore LN2-based automated sample stores (Cryo M42, Cryo M60, Cryo M60 for 250 mL cassettes and Ultra Low M60), not just the scenarios shown below.
Any configuration of these four BioStore models is in scope — not only the scenarios pictured here. Not covered: prices (no public price list was found, so this validator produces no price build-up or line-item quote), other labware and box types, other Azenta stores (ambient, −20 °C and −80 °C mechanical stores), software and LIMS options, power and LN2 supply, floor loading, service and regional terms.
8 scenarios in 4 groups · each group runs from the agent’s first attempt to the outcome · click one to switch
Illustrative agent statement. The validator receives the structured request below.
Proposing agent: AI sales agent (scripted)
Before sending, the agent must call Recorded run
What this means
Our assessment of what a typical rules-based CPQ (product rules, price rules and validation rules, as in Salesforce CPQ) does with the same cases. It is a judgment about effort and reliability, not a claim that nothing can be built.
| What the validator caught | Typical rules-based CPQ | Predictika |
|---|---|---|
| Capacity that depends on model and labware | Only with care Each model holds a different number of each labware type, so every pair needs its own capacity value and a rule that compares it with the quantity ordered. | The model, labware and quantity arrive together; the published capacity is looked up and compared on every call. |
| A temperature setpoint range that depends on the model | Only with care The Cryo stores operate at one temperature and the Ultra Low store over a range, so the allowed setpoint changes with the model and each needs its own rule. | Model and setpoint are checked together on every call, whichever order they arrive in. |
| Room dimensions that depend on the model | Only with workarounds Height, width and depth limits differ by model and come from the customer site, a different record than the product, so a rule has to reach across them. | The site dimensions travel with the configuration and are compared with the dimensions of the chosen model. |
| Stated capacity that must match the published figure | Only with care A stated figure is a free-entry field; a rule can compare it with the published capacity, but someone has to write and maintain that comparison. | The stated capacity travels with the configuration and is compared with the published capacity on every call. |
| Every problem returned in one answer | Only with workarounds Rules tend to fire as the order is saved or a step is completed, so a rep often fixes one error and meets the next. | All findings come back together, whatever order the choices arrived in. |
Every rule is checked on every call, in any order the choices arrive; one problem never hides another. Blocking findings stop the quote; approval findings need a person to sign off; everything else is computed.
| 1 | The quote needs room for 63,001 items; the selected store holds 63,000 of the selected labware. |
| 2 | The quote needs room for 123,489 items; the selected store holds 123,488 of the selected labware. |
| 3 | The quote needs room for 25,201 items; the selected store holds 25,200 of the selected labware. |
| 4 | The quote needs room for 49,393 items; the selected store holds 49,392 of the selected labware. |
| 5 | The quote asks for a setpoint of -19 degrees C; the Ultra Low store is set between -20 and -150 degrees C. |
| 6 | The quote asks for a setpoint of -151 degrees C; the Ultra Low store is set between -20 and -150 degrees C. |
| 7 | The quote asks for a setpoint of -189 degrees C; the Cryo store operates at -190 degrees C or colder. |
| 8 | The quote asks for a setpoint of -150 degrees C; the Cryo store operates at -190 degrees C or colder. |
| 9 | The quote needs room for 991 items; the selected store holds 990 of the selected labware. |
| 10 | No capacity is published for this labware in this store model; a person should confirm the labware and capacity with Azenta. |
| 11 | The quote asks for a setpoint of -190 degrees C; the Ultra Low store is set between -20 and -150 degrees C. |
| 12 | The room is 124 inches high; the selected store is 125 inches high. |
| 13 | The room is 107 inches high; the selected store is 108 inches high. |
| 14 | The space is 41 inches wide; the selected store is 42 inches wide. |
| 15 | The space is 61 inches deep; the selected store is 62 inches deep. |
| 16 | The space is 59 inches wide; the selected store is 60 inches wide. |
| 17 | The space is 78 inches deep; the selected store is 79 inches deep. |
| 18 | The room temperature is 64 degrees F; the store operates in a room between 65 and 80 degrees F. |
| 19 | The room temperature is 81 degrees F; the store operates in a room between 65 and 80 degrees F. |
| 20 | The room humidity is 51 percent; the store operates at no more than 50 percent relative humidity, non-condensing. |
| 21 | The quote asks for storage of 0 items; at least 1 is needed. |
| 22 | The quote states a capacity of 63,001 items; the published capacity of the selected store and labware is 63000. |
| 23 | The quote states a capacity of 62,999 items; the published capacity of the selected store and labware is 63000. |
| 24 | The quote needs room for 30,000 items; the selected store holds 25,200 of the selected labware. |
| 25 | The quote asks for a setpoint of -100 degrees C; the Cryo store operates at -190 degrees C or colder. |
| 26 | The room is 100 inches high; the selected store is 108 inches high. |
| 27 | The room temperature is 90 degrees F; the store operates in a room between 65 and 80 degrees F. |
| 28 | The room humidity is 60 percent; the store operates at no more than 50 percent relative humidity, non-condensing. |
The recorded result of the validator for each scenario on this page: the request the AI agent sent, the verdict, and every finding, in the order they appear above.
The agent's request. Please quote a BioStore Cryo M60 to hold 70,000 2 mL tubes at a setpoint of minus 190 degrees C, installed in a space 118 inches high, 80 wide and 100 deep, with the room at 72 degrees F and 40 percent humidity. The quote states a capacity of 63,000. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote a BioStore Cryo M60 to hold 60,000 2 mL tubes at a setpoint of minus 190 degrees C, installed in a space 130 inches high, 80 wide and 100 deep, with the room at 72 degrees F and 40 percent humidity. The quote states a capacity of 63,000. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote a BioStore Ultra Low M60 to hold 100,000 1 mL tubes at a setpoint of minus 170 degrees C, installed in a space 150 inches high, 70 wide and 90 deep, with the room at 84 degrees F and 45 percent humidity. The quote states a capacity of 123,488. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote a BioStore Ultra Low M60 to hold 100,000 1 mL tubes at a setpoint of minus 100 degrees C, installed in a space 150 inches high, 70 wide and 90 deep, with the room at 74 degrees F and 45 percent humidity. The quote states a capacity of 123,488. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote a BioStore Cryo M60 for 250 mL cassettes to hold 800 1 mL tubes at a setpoint of minus 190 degrees C, installed in a space 130 inches high, 62 wide and 85 deep, with the room at 70 degrees F and 35 percent humidity. It is ready to send.
Verdict: Needs approval.
The agent's request. Please quote a BioStore Cryo M60 for 250 mL cassettes to hold 800 250 mL cassettes at a setpoint of minus 190 degrees C, installed in a space 130 inches high, 62 wide and 85 deep, with the room at 70 degrees F and 35 percent humidity. The quote states a capacity of 990. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote a BioStore Cryo M42 to hold 49,392 1 mL tubes at a setpoint of minus 190 degrees C, installed in a space 110 inches high, 45 wide and 65 deep, with the room at 72 degrees F and 40 percent humidity. The quote states a capacity of 123,488. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote a BioStore Cryo M42 to hold 49,392 1 mL tubes at a setpoint of minus 190 degrees C, installed in a space 110 inches high, 45 wide and 65 deep, with the room at 72 degrees F and 40 percent humidity. The quote states a capacity of 49,392. It is ready to send.
Verdict: Allowed.
The text of the narrated walkthrough, step by step.