Predictika · Diagnostic X-ray imaging
An AI sales agent specifies a Carestream DRX-Compass radiography room for an imaging center: the tube, the tube positioner, the generator and the site power, the detector and how it connects, whether it sits in the wall stand, the heaviest patient and the tube heat capacity the quote states. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model of the public DRX-Compass component specification and returns a verdict, every reason and the figures it computed.
This validator models the DRX-Compass X-ray system from the public Carestream component specification (AK5184 Rev E), not just the scenarios shown below.
Any DRX-Compass room built from these components is in scope — not only the scenarios pictured here. Not covered: prices (Carestream publishes no list prices here, so this page shows no price build-up), the table and wall stand models, collimators, grids, workstation software and options, other Carestream systems such as the DRX-Evolution, and generator mA output (the specification lists the CGN-80 under two output columns). Assumptions: the Focus detectors and the 2530C are read as not tethered because the specification lists them as wireless only; the specification is a distributor-hosted copy of Carestream document AK5184 Rev E.
7 scenarios in 3 groups · each group runs from the agent’s first attempt to the outcome · click one to switch
Illustrative agent statement. The validator receives the structured request below.
Proposing agent: AI sales agent (scripted)
Before sending, the agent must call Recorded run
What this means
Our assessment of what a typical rules-based CPQ (product rules, price rules and validation rules, as in Salesforce CPQ) does with the same cases. It is a judgment about effort and reliability, not a claim that nothing can be built.
| What the validator caught | Typical rules-based CPQ | Predictika |
|---|---|---|
| Generators that fit only one site power supply | Only with workarounds Each generator is listed for one supply, so a rule has to read the generator on one line and the site power on another, for every pairing. | The generator and the site power are in one request and the match is checked with everything else. |
| A tube that can be used on only one positioner | Only with workarounds The restriction depends on the tube on one line and the positioner on another, and a miss is usually found at installation. | Both are in the request and the restriction comes back as its own finding. |
| Which detectors can be tethered or used in the wall stand | Only with care A product rule per detector, kept in step with the specification. | The detector selects its connection and use limits inside one model. |
| Limits that depend on the tube: kV and heat units | Only with care Each tube has its own voltage ceiling and heat capacity, so every exam value has to be compared per tube. | The tube selects its limits and every one is checked together. |
| Every problem returned in one answer | Only with workarounds Rules tend to fire as the quote is saved or a step is completed, so a person often fixes one error and meets the next. | All findings come back together, whatever order the choices arrived in. |
Every rule is checked on every call, in any order the choices arrive; one problem never hides another. Blocking findings stop the quote; approval findings need a person to sign off; everything else is computed.
| 1 | This is a low-speed tube, and the specification notes that the generator maximum kW output may be limited with a low-speed tube; a person should confirm the output meets the exam needs. |
| 2 | The Canon E7242X tube can be used only on the floor-mount tube stand, not the overhead tube crane. |
| 3 | The quoted generator is not listed for this site power supply. |
| 4 | The exams need 151 kV; this tube goes up to 150 kV. |
| 5 | The exams need 126 kV; this tube goes up to 125 kV. |
| 6 | The quote tethers this detector, but it is listed as wireless only. |
| 7 | The DRX-Plus 2530C detector is listed for tabletop use only, not for the wall stand. |
| 8 | The heaviest patient weighs 296 kg; the table takes at most 295 kg. |
| 9 | The quote states 400,000 heat units for the tube anode; the tube has 300,000. |
| 10 | The exams need 140 kV; this tube goes up to 125 kV. |
| 11 | The heaviest patient weighs 310 kg; the table takes at most 295 kg. |
| 12 | The quote states 300,000 heat units for the tube anode; the tube has 200,000. |
The recorded result of the validator for each scenario on this page: the request the AI agent sent, the verdict, and every finding, in the order they appear above.
The agent's request. Please quote a DRX-Compass room with a Canon E7242X tube on the overhead tube crane, a CGN-80 generator on three-phase 380/400/480 VAC site power and a DRX-Plus 3543C detector. The exams need up to 130 kV and the heaviest patient weighs 150 kg. The quote states 200,000 heat units for the tube anode. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote a DRX-Compass room with a Canon E7254FX tube on the overhead tube crane, a CGN-80 generator on three-phase 380/400/480 VAC site power and a DRX-Plus 3543C detector. The exams need up to 130 kV and the heaviest patient weighs 150 kg. The quote states 400,000 heat units for the tube anode. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote a DRX-Compass room with a Canon E7242X tube on the floor-mount tube stand, a CGN-80 generator on single-phase 200 to 260 VAC site power and a Focus 35C detector, tethered. The exams need up to 110 kV and the heaviest patient weighs 140 kg. The quote states 200,000 heat units for the tube anode. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote a DRX-Compass room with a Canon E7242X tube on the floor-mount tube stand, a CGF-40-1 generator on single-phase 200 to 260 VAC site power and a DRX-Plus 3543C detector, tethered. The exams need up to 110 kV and the heaviest patient weighs 140 kg. The quote states 200,000 heat units for the tube anode. It is ready to send.
Verdict: Needs approval.
The agent's request. Please quote a DRX-Compass room with a Canon E7252X tube on the floor-mount tube stand, a CGF-40-1 generator on single-phase 200 to 260 VAC site power and a DRX-Plus 3543C detector, tethered. The exams need up to 110 kV and the heaviest patient weighs 140 kg. The quote states 300,000 heat units for the tube anode. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote a DRX-Compass room with a Canon E7254FX tube on the floor-mount tube stand, a CGN-80 generator on three-phase 380/400/480 VAC site power and a DRX-Plus 2530C detector used in the wall stand. The exams need up to 120 kV and the heaviest patient weighs 310 kg. The quote states 400,000 heat units for the tube anode. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote a DRX-Compass room with a Canon E7254FX tube on the floor-mount tube stand, a CGN-80 generator on three-phase 380/400/480 VAC site power and a DRX-Plus 3543C detector used in the wall stand. The exams need up to 120 kV and the heaviest patient weighs 290 kg. The quote states 400,000 heat units for the tube anode. It is ready to send.
Verdict: Allowed.
The text of the narrated walkthrough, step by step.