An AI sales agent prepares an instrument order for a core facility or lab: the laser configuration (or a laser upgrade), the plate loader, and the site it will be installed in. Before the order goes out, the agent sends it to Predictika, which checks it against a declarative model of the published product documents and returns a verdict, every reason and the figures it computed.
This validator models orders for the Cytek Aurora analyzer, not just the scenarios shown below.
Any Aurora order that touches these rules is in scope — not only the scenarios pictured here. Not covered: prices (instruments and service plans are not published and no price list is public, so this validator produces no price build-up or line items), service plan terms, training, the Aurora CS sorter, the Northern Lights and other Cytek instruments, reagents and software.
7 scenarios in 3 groups · each group runs from the agent’s first attempt to the outcome · click one to switch
Illustrative agent statement. The validator receives the structured request below.
Proposing agent: AI sales agent (scripted)
Before sending, the agent must call Recorded run
What this means
Our assessment of what a typical rules-based CPQ (product rules, price rules and validation rules, as in Salesforce CPQ) does with the same cases. It is a judgment about effort and reliability, not a claim that nothing can be built.
| What the validator caught | Typical rules-based CPQ | Predictika |
|---|---|---|
| A laser upgrade that drops or repeats a laser | Only with care The valid paths depend on which lasers are already installed; every starting configuration needs its own rule. | Existing and target lasers arrive in one request and are compared on every call. |
| Plate formats a loader cannot run | Only with workarounds Which plate types each loader takes is a table that must be kept per loader. | Loader and plate type are checked together on every call. |
| Site power, temperature and humidity | Only with care Alternative voltage and current pairs plus numeric ranges have to be written as separate rules. | Each site figure is compared with its published range on every call. |
| Channel count against the lasers | Only with workarounds The channel count is a sum over the installed lasers that must be compared with a figure typed on the order. | The count is computed on every call and compared with the stated one. |
| Every problem returned in one answer | Only with workarounds Rules tend to fire as the order is saved or a step is completed, so a rep often fixes one error and meets the next. | All findings come back together, whatever order the choices arrived in. |
Every rule is checked on every call, in any order the choices arrive; one problem never hides another. Blocking findings stop the order; approval findings need a person to sign off; everything else is computed.
| 1 | The upgrade leaves out a laser the system already has; the Aurora upgrade paths only add the 561 nm yellow-green or the 355 nm ultraviolet laser to the base 405, 488 and 640 nm lasers. |
| 2 | The upgrade order has the same lasers as the existing system, so it adds no laser. |
| 3 | The plate or rack on the order cannot be run by the selected loader; the AMS takes 96-well plates only, the ASL takes 96-well plates, 96-deep well plates and 40-tube racks, and plates or racks need a loader. |
| 4 | The site supply is 120 V at 10 A; the Aurora needs 100 to 140 VAC at 15 A or 200 to 250 VAC at 10 A. |
| 5 | The site supply is 230 V at 5 A; the Aurora needs 100 to 140 VAC at 15 A or 200 to 250 VAC at 10 A. |
| 6 | The site supply is 170 V at 20 A; the Aurora needs 100 to 140 VAC at 15 A or 200 to 250 VAC at 10 A. |
| 7 | The site supply is 251 V at 10 A; the Aurora needs 100 to 140 VAC at 15 A or 200 to 250 VAC at 10 A. |
| 8 | The room temperature is 14 degrees C; the Aurora needs 15 to 28 degrees C (59 to 82.4 degrees F). |
| 9 | The room temperature is 29 degrees C; the Aurora needs 15 to 28 degrees C (59 to 82.4 degrees F). |
| 10 | The room humidity is 19 percent; the Aurora needs 20 to 85 percent relative humidity, non-condensing. |
| 11 | The room humidity is 86 percent; the Aurora needs 20 to 85 percent relative humidity, non-condensing. |
| 12 | The order states 54 fluorescence channels, but this laser configuration has 48 (16 violet, 14 blue and 8 red, plus 16 ultraviolet and 10 yellow-green when those lasers are installed). |
| 13 | The workspace on the order is 150 x 80 x 135 cm (width x depth x height), below the 165 x 76 x 132 cm Cytek recommends; confirm it with Cytek. |
| 14 | The room temperature is 30 degrees C; the Aurora needs 15 to 28 degrees C (59 to 82.4 degrees F). |
| 15 | The room humidity is 90 percent; the Aurora needs 20 to 85 percent relative humidity, non-condensing. |
| 16 | The order states 48 fluorescence channels, but this laser configuration has 64 (16 violet, 14 blue and 8 red, plus 16 ultraviolet and 10 yellow-green when those lasers are installed). |
The recorded result of the validator for each scenario on this page: the request the AI agent sent, the verdict, and every finding, in the order they appear above.
The agent's request. Please order a new 5-laser Cytek Aurora with the AMS loader running 40-tube racks, installed at a site with 120 V at 15 A, a room at 22 degrees C and 45 percent humidity. The order states 64 fluorescence channels. It is ready to send.
Verdict: Blocked.
The agent's request. Please order a new 5-laser Cytek Aurora with the ASL loader running 40-tube racks, installed at a site with 120 V at 15 A, a room at 22 degrees C and 45 percent humidity. The order states 64 fluorescence channels. It is ready to send.
Verdict: Allowed.
The agent's request. Please order a new 4-laser with the 355 nm ultraviolet laser Cytek Aurora with the no loader running manual tubes, installed at a site with 120 V at 10 A, a room at 30 degrees C and 45 percent humidity. The order states 54 fluorescence channels. It is ready to send.
Verdict: Blocked.
The agent's request. Please order a new 4-laser with the 355 nm ultraviolet laser Cytek Aurora with the no loader running manual tubes, installed at a site with 120 V at 15 A, a room at 30 degrees C and 45 percent humidity. The order states 54 fluorescence channels. It is ready to send.
Verdict: Blocked.
The agent's request. Please order a new 4-laser with the 355 nm ultraviolet laser Cytek Aurora with the no loader running manual tubes, installed at a site with 120 V at 15 A, a room at 24 degrees C and 45 percent humidity. The order states 54 fluorescence channels. It is ready to send.
Verdict: Allowed.
The agent's request. Please order a laser upgrade of an existing 4-laser yellow-green Cytek Aurora to the 4-laser with the 355 nm ultraviolet laser configuration with the no loader running manual tubes, installed at a site with 120 V at 15 A, a room at 22 degrees C and 45 percent humidity. The order states 64 fluorescence channels. It is ready to send.
Verdict: Blocked.
The agent's request. Please order a laser upgrade of an existing 4-laser yellow-green Cytek Aurora to the 5-laser configuration with the no loader running manual tubes, installed at a site with 120 V at 15 A, a room at 22 degrees C and 45 percent humidity. The order states 64 fluorescence channels. It is ready to send.
Verdict: Allowed.
The text of the narrated walkthrough, step by step.