Predictika · Anaesthesia workstation orders
An AI sales agent prepares an order for an Atlan A350/A350 XL anaesthesia workstation for a hospital operating-room project: gas version, gas cylinders, tidal volume range and a set of software, monitoring and connectivity options. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model of the product information Draeger publishes and returns a verdict, every reason and the figures it computed.
This validator models Atlan A350/A350 XL workstation orders, not just the scenarios shown below.
Any Atlan A350/A350 XL order that touches these rules is in scope — not only the scenarios pictured here. Not covered: prices (hospital pricing is project-based and not published, so this validator produces no price build-up or line-item quote), the difference between the A350 and the A350 XL, vaporizer counts, mounting versions and trolley sizes, ventilation-mode packaging, other Draeger products such as the Perseus A500, ventilators and ceiling supply units, and any clinical use. It is not a clinical tool.
7 scenarios in 3 groups · each group runs from the agent’s first attempt to the outcome · click one to switch
Illustrative agent statement. The validator receives the structured request below.
Proposing agent: AI sales agent (scripted)
Before sending, the agent must call Recorded run
What this means
Our assessment of what a typical rules-based CPQ (product rules, price rules and validation rules, as in Salesforce CPQ) does with the same cases. It is a judgment about effort and reliability, not a claim that nothing can be built.
| What the validator caught | Typical rules-based CPQ | Predictika |
|---|---|---|
| Options that exist only with a particular module or monitor | Only with care A product rule per option naming the module it depends on; it works, but every new option is another rule to maintain. | The options and the modules they depend on are in one request and every prerequisite is checked together. |
| Options that need a minimum software version | Only with care A rule per option against a software-version attribute, kept in step with each release. | The software version and the options travel in one request and are checked on every call. |
| A stated lowest tidal volume that must match the configuration | Only with workarounds The figure depends on whether an option is on the order, so it needs a derived value and a comparison rule. | The configuration's lowest tidal volume is computed and compared with the figure the order states. |
| A site-infrastructure option that needs a person to confirm | Only with workarounds An approval step that depends on one option but must hold the whole order until a person signs off. | The approval comes back with the other findings and the order is held until it is given. |
| Every problem returned in one answer | Only with workarounds Rules tend to fire as the order is saved or a step is completed, so a rep often fixes one error and meets the next. | All findings come back together, whatever order the choices arrived in. |
Every rule is checked on every call, in any order the choices arrive; one problem never hides another. Blocking findings stop the quote; approval findings need a person to sign off; everything else is computed.
| 1 | Advanced Gas Monitoring is available only with the integrated patient-gas measurement module; the quote has no such module. |
| 2 | Automatic xMAC monitoring is available only with the integrated patient-gas measurement module; the quote has no such module. |
| 3 | The O2 real gas test is available only with the integrated patient-gas measurement module; the quote has no such module. |
| 4 | Advanced Ventilation Monitoring is available only with Draeger Infinity Acute Care System (IACS) patient monitoring; the quote has none. |
| 5 | Lung Recruitment Maneuvers require software 2.0 or higher; the quote is on an earlier software version. |
| 6 | The Auto On function requires software 2.0 or higher; the quote is on an earlier software version. |
| 7 | Export and import of the machine configuration via USB requires software 2.0 or higher; the quote is on an earlier software version. |
| 8 | SmartPilot View software requires a medical-grade PC; the quote has none. |
| 9 | Connected Maintenance requires the Connectivity Package to establish connectivity; the quote has none. |
| 10 | Time and date synchronisation requires the Draeger Connectivity Converter hardware; the quote has none. |
| 11 | Data export and EMR integration require the Draeger Connectivity Converter hardware; the quote has none. |
| 12 | MobilePatientWatch requires compatible medical devices and additional IT infrastructure; a person should confirm the site has them. |
| 13 | A tidal volume of 8 mL is required, but this configuration delivers down to 10 mL (10 mL standard, 5 mL with Advanced neonatal support). |
| 14 | A tidal volume of 1,600 mL is required, but the maximum tidal volume is 1,500 mL. |
| 15 | The order states that the machine delivers down to 5 mL, but this configuration delivers down to 10 mL. |
| 16 | Nitrous oxide (N2O) is required but the 2-gas version supplies only O2 and air; the 3-gas version adds N2O. |
| 17 | 3 standing gas cylinders are quoted; the options are 1 or 2 standing cylinders. |
| 18 | 1 suspended gas cylinder are quoted; the options are 2 or 3 suspended cylinders. |
| 19 | A tidal volume of 7 mL is required, but this configuration delivers down to 10 mL (10 mL standard, 5 mL with Advanced neonatal support). |
The recorded result of the validator for each scenario on this page: the request the AI agent sent, the verdict, and every finding, in the order they appear above.
The agent's request. Please quote an Atlan A350/A350 XL anaesthesia workstation, 3-gas (O2/air/N2O) version, with 2 standing gas cylinders, the integrated patient-gas measurement module and IACS patient monitoring, with Lung Recruitment Maneuvers, Auto On, USB configuration transfer, SmartPilot View, Connected Maintenance and time synchronisation. The ventilator will be set to a tidal volume of 150 mL, and the order states that the machine delivers down to 10 mL. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote an Atlan A350/A350 XL anaesthesia workstation, 3-gas (O2/air/N2O) version, with 2 standing gas cylinders, the integrated patient-gas measurement module, IACS patient monitoring, software 2.0 and a medical-grade PC, with Lung Recruitment Maneuvers, Auto On, USB configuration transfer, SmartPilot View, Connected Maintenance and time synchronisation. The ventilator will be set to a tidal volume of 150 mL, and the order states that the machine delivers down to 10 mL. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote an Atlan A350/A350 XL anaesthesia workstation, 3-gas (O2/air/N2O) version, with 2 standing gas cylinders, the integrated patient-gas measurement module, IACS patient monitoring, software 2.0, a medical-grade PC, the Connectivity Package and the Connectivity Converter, with Lung Recruitment Maneuvers, Auto On, USB configuration transfer, SmartPilot View, Connected Maintenance and time synchronisation. The ventilator will be set to a tidal volume of 150 mL, and the order states that the machine delivers down to 10 mL. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote an Atlan A350/A350 XL anaesthesia workstation, 3-gas (O2/air/N2O) version, with 2 standing gas cylinders, software 2.0, with Advanced Gas Monitoring, xMAC monitoring, the O2 real gas test and Advanced Ventilation Monitoring. The ventilator will be set to a tidal volume of 150 mL, and the order states that the machine delivers down to 10 mL. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote an Atlan A350/A350 XL anaesthesia workstation, 3-gas (O2/air/N2O) version, with 2 standing gas cylinders, the integrated patient-gas measurement module, IACS patient monitoring and software 2.0, with Advanced Gas Monitoring, xMAC monitoring, the O2 real gas test and Advanced Ventilation Monitoring. The ventilator will be set to a tidal volume of 150 mL, and the order states that the machine delivers down to 10 mL. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote an Atlan A350/A350 XL anaesthesia workstation, 2-gas (O2/air) version, with 3 standing gas cylinders, the integrated patient-gas measurement module, IACS patient monitoring and software 2.0, with Advanced Gas Monitoring. The theatre needs nitrous oxide. The ventilator will be set to a tidal volume of 6 mL, and the order states that the machine delivers down to 5 mL. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote an Atlan A350/A350 XL anaesthesia workstation, 3-gas (O2/air/N2O) version, with 2 standing gas cylinders, the integrated patient-gas measurement module, IACS patient monitoring and software 2.0, with Advanced Gas Monitoring and Advanced neonatal support. The theatre needs nitrous oxide. The ventilator will be set to a tidal volume of 6 mL, and the order states that the machine delivers down to 5 mL. It is ready to send.
Verdict: Allowed.
The text of the narrated walkthrough, step by step.