Predictika · Transcatheter heart valve case orders
An AI sales agent prepares an order for a transcatheter heart valve case: the valve size chosen from the patient's CT annulus area, the delivery system, the introducer sheath and the vessel access. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model of the sizing and compatibility rules and returns a verdict, every reason and the figures it computed.
This validator models SAPIEN 3 and SAPIEN 3 Ultra valve case orders as described in the public instructions for use, not just the scenarios shown below.
Any SAPIEN 3 / SAPIEN 3 Ultra valve case order that touches these rules is in scope — not only the scenarios pictured here. Not covered: prices (hospital contract pricing is not public, so this validator produces no price build-up or line-item quote), other Edwards valves and monitoring products, clinical decisions, and delivery routes other than the table values applied here as stated.
8 scenarios in 3 groups · each group runs from the agent’s first attempt to the outcome · click one to switch
Illustrative agent statement. The validator receives the structured request below.
Proposing agent: AI sales agent (scripted)
Before sending, the agent must call Recorded run
What this means
Our assessment of what a typical rules-based CPQ (product rules, price rules and validation rules, as in Salesforce CPQ) does with the same cases. It is a judgment about effort and reliability, not a claim that nothing can be built.
| What the validator caught | Typical rules-based CPQ | Predictika |
|---|---|---|
| A valve size that must fit the measured annulus area | Only with care Each size has its own area range; a rule per size works, but the ranges overlap and have to be kept in step with the instructions for use. | The area is compared with the range of the chosen size on every call, and areas that fit two sizes are flagged for a person. |
| A delivery system that must match the valve size | Only with care A product rule per size pairing. It works, but every new size or model is another set of rules to maintain. | The valve and delivery system are in one request and the pairing is checked with every other rule. |
| A minimum vessel diameter that depends on the valve size | Only with workarounds The limit depends on the size on one line and a patient measurement on another, so it needs a cross-line rule that interacts with the rest. | Every choice and measurement is in one request, so the conflict is found whichever order they arrive in. |
| A sheath requirement that applies to one size only | Only with workarounds A rule that applies to a single size of one product against an accessory on another line, easy to miss in a generic rule set. | Size and sheath are checked together on every call. |
| Every problem returned in one answer | Only with workarounds Rules tend to fire as the order is saved or a step is completed, so a rep often fixes one error and meets the next. | All findings come back together, whatever order the choices arrived in. |
Every rule is checked on every call, in any order the choices arrive; one problem never hides another. Blocking findings stop the quote; approval findings need a person to sign off; everything else is computed.
| 1 | The CT annulus area is 272 mm2; the 20 mm valve is sized for 273 to 345 mm2. |
| 2 | The CT annulus area is 684 mm2; the 29 mm valve is sized for 540 to 683 mm2. |
| 3 | The CT annulus area is 500 mm2; the 23 mm valve is sized for 338 to 430 mm2. |
| 4 | The 26 mm valve needs the 26 mm COMMANDER delivery system; the order has the 23 mm one. |
| 5 | The 20 mm valve needs the 20 mm COMMANDER delivery system; the order has the 29 mm one. |
| 6 | The minimum vessel diameter for the 26 mm valve is 5.5 mm; the vessel diameter on the order is 5.4 mm. |
| 7 | The minimum vessel diameter for the 29 mm valve is 6.0 mm; the vessel diameter on the order is 5.9 mm. |
| 8 | The 29 mm system uses a 16F introducer sheath set or equivalent; the order has a 14F set. |
| 9 | The CT annulus area of 340 mm2 falls in the range of two valve sizes; a person should confirm the size choice. |
| 10 | The CT annulus area of 430 mm2 falls in the range of two valve sizes; a person should confirm the size choice. |
| 11 | The CT annulus area of 545 mm2 falls in the range of two valve sizes; a person should confirm the size choice. |
| 12 | The CT annulus area is 547 mm2; the 26 mm valve is sized for 430 to 546 mm2. |
| 13 | The 20 mm valve needs the 20 mm COMMANDER delivery system; the order has the 23 mm one. |
| 14 | The CT annulus area is 500 mm2; the 29 mm valve is sized for 540 to 683 mm2. |
| 15 | The 29 mm valve needs the 29 mm COMMANDER delivery system; the order has the 20 mm one. |
| 16 | The minimum vessel diameter for the 29 mm valve is 6.0 mm; the vessel diameter on the order is 5.0 mm. |
| 17 | The CT annulus area of 338 mm2 falls in the range of two valve sizes; a person should confirm the size choice. |
| 18 | The CT annulus area is 337 mm2; the 23 mm valve is sized for 338 to 430 mm2. |
The recorded result of the validator for each scenario on this page: the request the AI agent sent, the verdict, and every finding, in the order they appear above.
The agent's request. Please order a 23 mm SAPIEN 3 valve with the 26 mm COMMANDER delivery system for a patient with a CT annulus area of 480 mm2 and a vessel diameter of 5.2 mm. It is ready to send.
Verdict: Blocked.
The agent's request. Please order a 26 mm SAPIEN 3 valve with the 26 mm COMMANDER delivery system for a patient with a CT annulus area of 480 mm2 and a vessel diameter of 5.6 mm. It is ready to send.
Verdict: Allowed.
The agent's request. Please order a 29 mm SAPIEN 3 valve with the 26 mm COMMANDER delivery system and a 14F introducer sheath set for a patient with a CT annulus area of 620 mm2 and a vessel diameter of 5.8 mm. It is ready to send.
Verdict: Blocked.
The agent's request. Please order a 29 mm SAPIEN 3 valve with the 29 mm COMMANDER delivery system and a 14F introducer sheath set for a patient with a CT annulus area of 620 mm2 and a vessel diameter of 6.2 mm. It is ready to send.
Verdict: Blocked.
The agent's request. Please order a 29 mm SAPIEN 3 valve with the 29 mm COMMANDER delivery system and a 16F introducer sheath set for a patient with a CT annulus area of 620 mm2 and a vessel diameter of 6.2 mm. It is ready to send.
Verdict: Allowed.
The agent's request. Please order a 20 mm SAPIEN 3 valve with the 23 mm COMMANDER delivery system for a patient with a CT annulus area of 340 mm2 and a vessel diameter of 5.5 mm. It is ready to send.
Verdict: Blocked.
The agent's request. Please order a 20 mm SAPIEN 3 valve with the 20 mm COMMANDER delivery system for a patient with a CT annulus area of 340 mm2 and a vessel diameter of 5.5 mm. It is ready to send.
Verdict: Needs approval.
The agent's request. Please order a 23 mm SAPIEN 3 valve with the 23 mm COMMANDER delivery system for a patient with a CT annulus area of 380 mm2 and a vessel diameter of 5.5 mm. It is ready to send.
Verdict: Allowed.
The text of the narrated walkthrough, step by step.