An AI sales agent prepares an order for a Proline Promag P 500 electromagnetic flowmeter: sensor size and liner, process connection, transmitter electronics, outputs and inputs, housing and cable. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model of the published ordering and application limits and returns a verdict, every reason and the figures it computed.
This validator models orders for the Proline Promag P 500 electromagnetic flowmeter, not just the scenarios shown below.
Any Promag P 500 order that touches these rules is in scope — not only the scenarios pictured here. Not covered: prices (Endress+Hauser does not publish list prices, so this validator produces no price build-up or line-item quote), the full order code and Product Configurator, hazardous-area approvals, electrode materials and chemical compatibility with the medium, pressure derating with temperature, ASME and JIS pressure ratings, vacuum limits, AS flanges, the cable length versus conductivity chart, application packages, accessories, and other Proline families.
7 scenarios in 3 groups · each group runs from the agent’s first attempt to the outcome · click one to switch
Illustrative agent statement. The validator receives the structured request below.
Proposing agent: AI sales agent (scripted)
Before sending, the agent must call Recorded run
What this means
Our assessment of what a typical rules-based CPQ (product rules, price rules and validation rules, as in Salesforce CPQ) does with the same cases. It is a judgment about effort and reliability, not a claim that nothing can be built.
| What the validator caught | Typical rules-based CPQ | Predictika |
|---|---|---|
| Outputs and inputs that are not offered together | Only with workarounds Which options exist for outputs 2, 3 and 4 depends on output 1, and the double pulse pairing and the output 4 limit add more; that is a table of excluded combinations across four choices that has to be built and kept in step by hand. | All four choices arrive in one request and are checked against the published combinations, whichever order they were made in. |
| A liner that cannot take the medium temperature or the line size | Only with care Each liner has its own temperature range and size range; a rule per liner works, but every new liner or limit is another rule to maintain. | Liner, diameter and the medium's temperature range are in one request and are compared on every call. |
| Ambient limits that depend on a different choice | Only with workarounds The sensor's lowest ambient temperature depends on the flange material, and the transmitter's on an option, while the two sit in different places; each needs its own cross-line rule. | Both locations are separate fields in the request; each is compared with the limit that follows from the choices actually made. |
| Cable length that depends on the electronics | Only with care The allowed cable is 300 m for one transmitter and 200 m for the other, so the rule needs the electronics choice and the length together. | The electronics and the length travel together and the limit is computed on every call. |
| Stated temperature rating that must match the liner | Only with care A stated figure is a free-entry field; a rule can compare it with the liner's rating, but someone has to write and maintain that comparison. | The stated figure travels with the order and is compared with the computed rating on every call. |
| Every problem returned in one answer | Only with workarounds Rules tend to fire as the order is saved or a step is completed, so a rep often fixes one error and meets the next. | All findings come back together, whatever order the choices arrived in. |
Every rule is checked on every call, in any order the choices arrive; one problem never hides another. Blocking findings stop the quote; approval findings need a person to sign off; everything else is computed.
| 1 | Output/input 4 is available only for the Proline 500 - digital transmitter. |
| 2 | This transmitter housing is not offered with the selected electronics; the digital transmitter comes in aluminum or polycarbonate, the standard transmitter in aluminum or cast stainless steel. |
| 3 | The option chosen for output/input 2 is not available with the option chosen for output/input 1. |
| 4 | The option chosen for output/input 3 is not available with the option chosen for output/input 1. |
| 5 | With a double pulse output on output/input 2, only the double pulse output (slave) is available for output/input 3. |
| 6 | Output/input 4 does not offer the Ex i passive current output, the double pulse output or the Ex i passive pulse/frequency/switch output. |
| 7 | A PFA liner is available for DN 25 to 200 only; the quote has DN 15, which needs a PTFE liner. |
| 8 | A PFA liner is available for DN 25 to 200 only; the quote has DN 250, which needs a PTFE liner. |
| 9 | The medium reaches 140 degrees C, but the selected liner is rated to 130 degrees C. |
| 10 | The medium reaches 160 degrees C, but the selected liner is rated to 150 degrees C. |
| 11 | The medium reaches 190 degrees C, but the selected liner is rated to 180 degrees C. |
| 12 | The medium falls to -30 degrees C, but the selected liner is rated down to -20 degrees C. |
| 13 | The medium conductivity of 4 microsiemens per cm is below the 5 microsiemens per cm minimum for liquids in general. |
| 14 | The transmitter must work down to -45 degrees C; the standard transmitter is rated to -40 degrees C and the ambient temperature option (-50 degrees C) is not on the quote. |
| 15 | The transmitter will see -45 degrees C; the local display is specified from -20 degrees C and its readability may be impaired below that. |
| 16 | The transmitter must work down to -55 degrees C, below the -50 degrees C of the ambient temperature option. |
| 17 | The transmitter will see -55 degrees C; the local display is specified from -20 degrees C and its readability may be impaired below that. |
| 18 | The sensor must work down to -15 degrees C; with the selected process connection material it is rated down to -10 degrees C. |
| 19 | There is no ASME B16.5 Class 150 flange for DN 65. |
| 20 | The operating pressure of 20 bar is above the 16 bar nominal rating of the EN 1,092-1 flange for this nominal diameter. |
| 21 | The operating pressure of 12 bar is above the 10 bar nominal rating of the EN 1,092-1 flange for this nominal diameter. |
| 22 | The connecting cable between sensor and transmitter is 250 m; this transmitter allows at most 200 m. |
| 23 | The connecting cable between sensor and transmitter is 350 m; this transmitter allows at most 300 m. |
| 24 | The quote states a medium temperature rating of 150 degrees C, but the selected liner is rated to 130 degrees C. |
| 25 | DN 20 is not a nominal diameter of the Promag P sensor (DN 15, 25, 32, 40, 50, 65, 80, 100, 125, 150, 200, 250, 300, 350, 400, 450, 500 or 600). |
The recorded result of the validator for each scenario on this page: the request the AI agent sent, the verdict, and every finding, in the order they appear above.
The agent's request. Please quote a Proline Promag P 500 with a DN 100 sensor, PTFE liner and EN 1092-1 carbon steel flanges, the Proline 500 transmitter with a housing made of aluminum, output/input 1: 4-20 mA HART Ex i passive, output/input 2: 4-20 mA, output/input 4: relay, and a 250 m connecting cable. The liquid is 800 µS/cm, runs from -10 to 90 °C at 10 bar; the sensor sees down to -5 °C and the transmitter down to -10 °C. The quote states that the liner is rated to 130 °C. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote a Proline Promag P 500 with a DN 100 sensor, PTFE liner and EN 1092-1 carbon steel flanges, the Proline 500 - digital transmitter with a housing made of aluminum, output/input 1: 4-20 mA HART Ex i passive, output/input 2: 4-20 mA Ex i passive, output/input 3: 4-20 mA Ex i passive, and a 250 m connecting cable. The liquid is 800 µS/cm, runs from -10 to 90 °C at 10 bar; the sensor sees down to -5 °C and the transmitter down to -10 °C. The quote states that the liner is rated to 130 °C. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote a Proline Promag P 500 with a DN 200 sensor, PFA liner and EN 1092-1 carbon steel flanges, the Proline 500 transmitter with a housing made of aluminum, output/input 1: 4-20 mA HART, output/input 2: pulse/frequency/switch, and a 40 m connecting cable. The liquid is 300 µS/cm, runs from 20 to 160 °C at 9 bar; the sensor sees down to 0 °C and the transmitter down to 0 °C. The quote states that the liner is rated to 160 °C. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote a Proline Promag P 500 with a DN 200 sensor, high-temperature PFA liner and EN 1092-1 carbon steel flanges, the Proline 500 transmitter with a housing made of aluminum, output/input 1: 4-20 mA HART, output/input 2: pulse/frequency/switch, and a 40 m connecting cable. The liquid is 300 µS/cm, runs from 20 to 160 °C at 9 bar; the sensor sees down to 0 °C and the transmitter down to 0 °C. The quote states that the liner is rated to 180 °C. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote a Proline Promag P 500 with a DN 65 sensor, PTFE liner and ASME B16.5 Class 150 carbon steel flanges, the Proline 500 transmitter with a housing made of aluminum, output/input 1: 4-20 mA HART, output/input 2: pulse/frequency/switch, and a 30 m connecting cable. The liquid is 500 µS/cm, runs from -5 to 60 °C at 10 bar; the sensor sees down to -20 °C and the transmitter down to -45 °C. The quote states that the liner is rated to 130 °C. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote a Proline Promag P 500 with a DN 65 sensor, PTFE liner and EN 1092-1 stainless steel flanges, the Proline 500 transmitter with a housing made of aluminum, output/input 1: 4-20 mA HART, output/input 2: pulse/frequency/switch, and a 30 m connecting cable, plus the -50 °C ambient temperature option. The liquid is 500 µS/cm, runs from -5 to 60 °C at 10 bar; the sensor sees down to -20 °C and the transmitter down to -45 °C. The quote states that the liner is rated to 130 °C. It is ready to send.
Verdict: Needs approval.
The agent's request. Please quote a Proline Promag P 500 with a DN 65 sensor, PTFE liner and EN 1092-1 stainless steel flanges, the Proline 500 transmitter with a housing made of aluminum, output/input 1: 4-20 mA HART, output/input 2: pulse/frequency/switch, and a 30 m connecting cable. The liquid is 500 µS/cm, runs from -5 to 60 °C at 10 bar; the sensor sees down to -20 °C and the transmitter down to -15 °C. The quote states that the liner is rated to 130 °C. It is ready to send.
Verdict: Allowed.
The text of the narrated walkthrough, step by step.