An AI sales agent prepares a hospital order for an Airvo 3 high flow system with a patient interface and size, a flow, a target temperature, an oxygen setting and the planned use of the tube and chamber kit, and states the maximum flow for the interface. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model of the specifications in the published Airvo 3 User Manual and returns a verdict, every reason and the figures it computed.
This validator models Airvo 3 high flow therapy orders as described in the public Airvo 3 User Manual, not just the scenarios shown below.
Any Airvo 3 high flow order that touches these rules is in scope — not only the scenarios pictured here. Not covered: prices (Fisher & Paykel Healthcare prices are set per hospital and distributor agreement and are not public, so this validator produces no price build-up or line-item quote), part numbers and bundles, the Optiflow+ Duet interface, the Airvo 2 system, the noninvasive ventilation modes and other Fisher & Paykel products. Assumptions: the listed flow range for each interface size is read as the allowed range for the order, and the 31 degrees C face mask guidance is treated as a point for a person to confirm. The model reflects the manual and regulatory summary read in October 2026.
7 scenarios in 3 groups · each group runs from the agent’s first attempt to the outcome · click one to switch
Illustrative agent statement. The validator receives the structured request below.
Proposing agent: AI sales agent (scripted)
Before sending, the agent must call Recorded run
What this means
Our assessment of what a typical rules-based CPQ (product rules, price rules and validation rules, as in Salesforce CPQ) does with the same cases. It is a judgment about effort and reliability, not a claim that nothing can be built.
| What the validator caught | Typical rules-based CPQ | Predictika |
|---|---|---|
| A flow range that depends on the interface and its size | Only with care Each interface and size pair has its own range, so the rules need a table of pairs and a check that reads it; each new size is another row to maintain. | The interface, size and flow are in one request; the listed range is looked up and compared with the flow on every call. |
| A size that exists only for some interfaces | Easy A compatibility rule per interface; routine in a product-rules CPQ. | The same rule, checked together with all the others. |
| A bypassed airway that limits temperature and flow together | Only with workarounds The limit depends on a patient flag, the temperature and the flow at once, which takes a cross-line rule that interacts with the range rules. | The flag, temperature and flow are in one request and the limit is checked with every other rule. |
| A stated maximum flow that disagrees with the interface | Only with workarounds The listed maximum is derived from the interface and size; checking a stated figure against it takes extra formulas that interact with the rest of the rules. | The stated maximum flow is a field in the request and is compared with the listed maximum. |
| Every problem returned in one answer | Only with workarounds Rules tend to fire as the order is saved or a step is completed, so a rep often fixes one error and meets the next. | All findings come back together, whatever order the choices arrived in. |
Every rule is checked on every call, in any order the choices arrive; one problem never hides another. Blocking findings stop the quote; approval findings need a person to sign off; everything else is computed.
| 1 | The order sets a flow of 61 liters a minute; this interface and size are listed for 10 to 60 liters a minute. |
| 2 | The order sets a flow of 9 liters a minute; this interface and size are listed for 10 to 60 liters a minute. |
| 3 | The order sets a flow of 71 liters a minute; this interface and size are listed for 10 to 70 liters a minute. |
| 4 | The selected size is not listed for this patient interface; Optiflow Junior 2 comes in sizes M, L, XL and XXL, Optiflow+ and Optiflow 3S in S, M and L, and the tracheostomy interface and mask adapter have no size. |
| 5 | The order sets a flow of 8 liters a minute; this interface and size are listed for 2 to 7 liters a minute. |
| 6 | The order sets a flow of 26 liters a minute; this interface and size are listed for 2 to 25 liters a minute. |
| 7 | The order sets a flow of 9 liters a minute; this interface and size are listed for 10 to 50 liters a minute. |
| 8 | For a patient with a bypassed airway the Airvo 3 is classified as a Category 1 humidifier only at 37 degrees C and 10 to 60 liters a minute; the order sets 34 degrees C and 40 liters a minute. |
| 9 | For a patient with a bypassed airway the Airvo 3 is classified as a Category 1 humidifier only at 37 degrees C and 10 to 60 liters a minute; the order sets 37 degrees C and 61 liters a minute. |
| 10 | Patients using face masks may find high temperatures uncomfortable and a target temperature of 31 degrees C is suggested; a person should confirm the 32 degrees C setting. |
| 11 | Patients using face masks may find high temperatures uncomfortable and a target temperature of 31 degrees C is suggested; a person should confirm the 37 degrees C setting. |
| 12 | The Airvo 3 target temperature runs from 31 to 37 degrees C; the order sets 30 degrees C. |
| 13 | The Airvo 3 target temperature runs from 31 to 37 degrees C; the order sets 38 degrees C. |
| 14 | The oxygen setting runs from 21 to 100 percent; the order sets 101 percent. |
| 15 | The oxygen setting runs from 21 to 100 percent; the order sets 20 percent. |
| 16 | An AirSpiral tube and chamber kit is listed for up to 14 days, or 7 days when using a nebulizer; the order plans 15 days and the limit here is 14. |
| 17 | An AirSpiral tube and chamber kit is listed for up to 14 days, or 7 days when using a nebulizer; the order plans 8 days and the limit here is 7. |
| 18 | An AirSpiral tube and chamber kit is listed for 1 patient; the order has 2 patients on one kit. |
| 19 | The order states a maximum flow of 60 liters a minute for this interface and size, but the listed maximum is 70 liters a minute. |
The recorded result of the validator for each scenario on this page: the request the AI agent sent, the verdict, and every finding, in the order they appear above.
The agent's request. Please prepare a hospital order for an Airvo 3 system with an Optiflow Junior 2 interface in size M, set to 12 liters a minute at 37 degrees C and 30 percent oxygen. The AirSpiral kit is planned for 10 days of use with a nebulizer, for 1 patient. The order states a maximum flow of 20 liters a minute for this interface. It is ready to send.
Verdict: Blocked.
The agent's request. Please prepare a hospital order for an Airvo 3 system with an Optiflow Junior 2 interface in size L, set to 12 liters a minute at 37 degrees C and 30 percent oxygen. The AirSpiral kit is planned for 7 days of use with a nebulizer, for 1 patient. The order states a maximum flow of 20 liters a minute for this interface. It is ready to send.
Verdict: Allowed.
The agent's request. Please prepare a hospital order for an Airvo 3 system with an Optiflow+ tracheostomy interface interface, set to 65 liters a minute at 34 degrees C and 105 percent oxygen, for a patient with a bypassed airway. The AirSpiral kit is planned for 7 days of use, for 1 patient. It is ready to send.
Verdict: Blocked.
The agent's request. Please prepare a hospital order for an Airvo 3 system with an Optiflow+ tracheostomy interface interface, set to 65 liters a minute at 37 degrees C and 60 percent oxygen, for a patient with a bypassed airway. The AirSpiral kit is planned for 7 days of use, for 1 patient. It is ready to send.
Verdict: Blocked.
The agent's request. Please prepare a hospital order for an Airvo 3 system with an Optiflow+ tracheostomy interface interface, set to 50 liters a minute at 37 degrees C and 60 percent oxygen, for a patient with a bypassed airway. The AirSpiral kit is planned for 7 days of use, for 1 patient. The order states a maximum flow of 60 liters a minute for this interface. It is ready to send.
Verdict: Allowed.
The agent's request. Please prepare a hospital order for an Airvo 3 system with an Optiflow+ mask adapter interface, set to 40 liters a minute at 37 degrees C and 60 percent oxygen. The AirSpiral kit is planned for 7 days of use, for 1 patient. The order states a maximum flow of 60 liters a minute for this interface. It is ready to send.
Verdict: Needs approval.
The agent's request. Please prepare a hospital order for an Airvo 3 system with an Optiflow+ mask adapter interface, set to 40 liters a minute at 31 degrees C and 60 percent oxygen. The AirSpiral kit is planned for 7 days of use, for 1 patient. The order states a maximum flow of 60 liters a minute for this interface. It is ready to send.
Verdict: Allowed.
The text of the narrated walkthrough, step by step.