Predictika · Electrosurgery unit orders
An AI sales agent prepares an order for KLS Martin maXium smart Q and maXium smart E electrosurgical units for a hospital: which variant, the software options for the current types the surgeons use, SWAP Mode, footswitches, the Beam argon flow, cart accessories, the site supply and the mains load it states. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model of what the public KLS Martin brochure and web shop say and returns a verdict, every reason and the figures it computed.
This validator models orders for new maXium smart Q and E units, all configured identically, based on the public KLS Martin brochure 91-350-87-02 and web shop article pages, not just the scenarios shown below.
Any order for these units that touches these rules is in scope — not only the scenarios pictured here. Not covered: prices (KLS Martin does not publish prices, so this page shows no price build-up), the maXium smart C, OR tables and lights, implants, and anything the brochure does not state.
7 scenarios in 3 groups · each group runs from the agent’s first attempt to the outcome · click one to switch
Illustrative agent statement. The validator receives the structured request below.
Proposing agent: AI sales agent (scripted)
Before sending, the agent must call Recorded run
What this means
Our assessment of what a typical rules-based CPQ (product rules, price rules and validation rules, as in Salesforce CPQ) does with the same cases. It is a judgment about effort and reliability, not a claim that nothing can be built.
| What the validator caught | Typical rules-based CPQ | Predictika |
|---|---|---|
| Options that depend on the current types in use | Only with care Each software option unlocks specific current types, so a rule per option must read what the surgeons will use. | The currents in use and the options on the order are checked together on every call. |
| Sockets that depend on the variant | Only with care The number of monopolar and bipolar sockets and the international socket differ by variant, so each variant needs its own rule. | Instruments needed are compared with the variant sockets in one place. |
| A stated total that must equal a sum | Only with care The mains load the order states must be compared with a sum over the units ordered, which rules must maintain. | The stated load is a field in the request and is compared with the computed sum. |
| Site conditions | Easy Ranges for mains, temperature and humidity are simple validation rules. | Checked with the rest, in the same answer. |
| Every problem returned in one answer | Only with workarounds Rules tend to fire as the order is saved or a step is completed, so a person often fixes one error and meets the next. | All findings come back together, whatever order the choices arrived in. |
Every rule is checked on every call, in any order the choices arrive; one problem never hides another. Blocking findings stop the order; approval findings need a person to sign off; everything else is computed.
| 1 | Bipolar sockets on this variant: 1; bipolar instruments the order needs connected at once: 2. |
| 2 | Monopolar sockets on this variant: 1; monopolar instruments the order needs connected at once: 2. |
| 3 | Monopolar sockets on this variant: 2; monopolar instruments the order needs connected at once: 3. |
| 4 | The maXium smart E MBB has only KLS Martin (m-version) sockets; international (i-version) instruments connect to the M2 socket of the Q (MMBB) and the E MMB. |
| 5 | The Polyp, Papilla and Mucosect currents need the GI software option (80-094-06-04), which must be ordered separately; it is not on the order. |
| 6 | The Seal Safe and Bipolate currents need the SealSafe IQ software option (80-094-05-04), which must be ordered separately; it is not on the order. |
| 7 | The Duo Prep Plus and Duo Spray currents need the Duo currents software option (80-094-01-04), which must be ordered separately; it is not on the order. |
| 8 | The Duo currents use both monopolar outputs at once; the E MBB has one monopolar socket, so a person should confirm the use. |
| 9 | The URO Bipolar Blend Cut, Pure Cut and Coag currents need the bipolar resection (TUR) software option (80-094-04-04), which must be ordered separately; it is not on the order. |
| 10 | The Neptun current needs the Neptun software option (80-094-02-04), which must be ordered separately; it is not on the order. |
| 11 | Switching between predefined programs from the sterile area is the SWAP Mode software option (80-094-00-04), which must be ordered separately; the order asks for 3 SWAP programs without it. |
| 12 | SWAP Mode switches between up to five predefined programs; the order asks for 6. |
| 13 | The dual-pedal footswitch with SWAP button (80-830-21-04) is on the order without the SWAP Mode option (80-094-00-04), so a person should confirm it. |
| 14 | The enhanced footswitch option (80-094-07-04) allows two dual-pedal footswitches; the order has 2 per unit without it, so a person should confirm it. |
| 15 | The maXium smart Beam argon flow is 0.1 to 12 l/min; the order asks for 13 l/min. |
| 16 | An argon flow is requested but no maXium smart Beam is on the order, so a person should confirm the customer has one. |
| 17 | The standard rail, separating transformer and gas bottle holder are described as for the maXium smart Cart; the order has no Cart, so a person should confirm the customer has one. |
| 18 | The unit takes 100 V to 240 V with a tolerance of 10 percent (90 V to 264 V) at 50/60 Hz; the site supply is 85 V. |
| 19 | The unit takes 100 V to 240 V with a tolerance of 10 percent (90 V to 264 V) at 50/60 Hz; the site supply is 265 V. |
| 20 | The unit operates at an ambient temperature of +10 to +40 °C; the room is at 8 °C. |
| 21 | The unit operates at an ambient temperature of +10 to +40 °C; the room is at 41 °C. |
| 22 | The unit operates at a relative humidity of 30 to 75 percent; the room is at 80 percent. |
| 23 | The order states a mains load of 1000 VA at maximum power; 2 units at 650 VA each add up to 1300 VA. |
| 24 | An order needs at least 1 unit; the order has 0. |
| 25 | The maXium smart Beam argon flow is 0.1 to 12 l/min; the order asks for 14 l/min. |
The recorded result of the validator for each scenario on this page: the request the AI agent sent, the verdict, and every finding, in the order they appear above.
The agent's request. Please prepare an order for 4 x maXium smart Q (MMBB, 80-043-20-04), each connected to 2 monopolar and 2 bipolar instruments. Software options: none. The surgeons will use GI currents (Polyp, Papilla, Mucosect) and vessel sealing (Seal Safe, Bipolate). Also: 1 dual-pedal footswitch per unit. The site supply is 230 V, the room is at 22 °C and 50 percent humidity. The order states a total mains load of 2,600 VA at maximum power. It is ready to send.
Verdict: Blocked.
The agent's request. Please prepare an order for 4 x maXium smart Q (MMBB, 80-043-20-04), each connected to 2 monopolar and 2 bipolar instruments. Software options: GI currents, SealSafe IQ. The surgeons will use GI currents (Polyp, Papilla, Mucosect) and vessel sealing (Seal Safe, Bipolate). Also: 1 dual-pedal footswitch per unit. The site supply is 230 V, the room is at 22 °C and 50 percent humidity. The order states a total mains load of 2,600 VA at maximum power. It is ready to send.
Verdict: Allowed.
The agent's request. Please prepare an order for 3 x maXium smart E MBB (80-043-12-04), each connected to 2 monopolar and 1 bipolar instrument. Software options: none. The surgeons will use bipolar resection currents and international instruments. The site supply is 230 V, the room is at 22 °C and 50 percent humidity. The order states a total mains load of 1,950 VA at maximum power. It is ready to send.
Verdict: Blocked.
The agent's request. Please prepare an order for 3 x maXium smart E MMB (80-043-11-04), each connected to 2 monopolar and 1 bipolar instrument. Software options: none. The surgeons will use bipolar resection currents and international instruments. The site supply is 230 V, the room is at 22 °C and 50 percent humidity. The order states a total mains load of 1,950 VA at maximum power. It is ready to send.
Verdict: Blocked.
The agent's request. Please prepare an order for 3 x maXium smart E MMB (80-043-11-04), each connected to 2 monopolar and 1 bipolar instrument. Software options: bipolar TUR. The surgeons will use bipolar resection currents and international instruments. The site supply is 230 V, the room is at 22 °C and 50 percent humidity. The order states a total mains load of 1,950 VA at maximum power. It is ready to send.
Verdict: Allowed.
The agent's request. Please prepare an order for 5 x maXium smart E MMB (80-043-11-04), each connected to 2 monopolar and 1 bipolar instrument. Software options: none. The surgeons will use the standard currents. Also: 5 maXium smart Beam at an argon flow of 15 l/min. The site supply is 230 V, the room is at 22 °C and 50 percent humidity. The order states a total mains load of 3,000 VA at maximum power. It is ready to send.
Verdict: Blocked.
The agent's request. Please prepare an order for 5 x maXium smart E MMB (80-043-11-04), each connected to 2 monopolar and 1 bipolar instrument. Software options: none. The surgeons will use the standard currents. Also: 5 maXium smart Beam at an argon flow of 10 l/min. The site supply is 230 V, the room is at 22 °C and 50 percent humidity. The order states a total mains load of 3,250 VA at maximum power. It is ready to send.
Verdict: Allowed.
The text of the narrated walkthrough, step by step.