An AI sales agent prepares an order for a Kewaunee Supreme Air Venturi fume hood from Kewaunee's public guide specification, with the series, face velocity, sash opening, hood length, liner, work surface, duct collar, the chemicals to be used and any fire suppression the lab needs. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model of the published specification and returns a verdict, every reason and the figures it computed.

Scope: Seventeen Kewaunee Supreme Air Venturi hood series: face velocity, sash opening, hood length, special-purpose materials, duct collars, fire suppression and liner chemical resistance. See everything this validator covers

This validator models orders for the 17 Supreme Air Venturi hood series in Kewaunee's public guide specification 11 53 13 (V05, V06, V07, V10, V11, V15, V16, V25, V26, V30, V36, V40, V45, V65, V66, V67 and V90), not just the scenarios shown below.

  • Performance — the face velocities the specification lists (50, 60, 80, 100 and 120 FPM), the lowest velocity each series is listed for, and the sash openings each series offers.
  • Size and counts — the hood lengths the specification names (4, 5, 6, 8, 10 and 12 ft), the duct collars a length needs (one up to 6 ft, two above) and the fire suppression units it needs (one up to 8 ft, two above); a stated count that differs is rejected, and fire suppression is checked only when the lab says it needs it.
  • Materials and chemicals — the perchloric acid hood (V45) must have 316L stainless steel liner, work surface and duct collar; the isotope hood (V40) the same in 304L; and for 49 chemicals the specification's liner resistance table decides: a D rating blocks the order and a C rating needs a person to confirm the liner.

Any order for these series that touches these rules is in scope — not only the scenarios pictured here. Not covered: prices (this validator produces no price build-up or line-item quote), other hood families, the chemical resistance of work surfaces and duct collars, fire suppression and duct collar options beyond the counts above, and accessories.

The AI sales agentBuilds the hood order from the conversation and says it is ready.
Predictika validatorChecks every rule at once and returns the verdict, each finding, and the computed figures.

6 scenarios in 3 groups · each group runs from the agent’s first attempt to the outcome · click one to switch

What the agent proposes

The AI sales agent says

Illustrative agent statement. The validator receives the structured request below.

Proposing agent: AI sales agent (scripted)

Before sending, the agent must call Recorded run

Predictika validator


      
Waiting for the call.

Where rules-based CPQ gets it wrong

Click to expand

Our assessment of what a typical rules-based CPQ (product rules, price rules and validation rules, as in Salesforce CPQ) does with the same cases. It is a judgment about effort and reliability, not a claim that nothing can be built.

What the validator caughtTypical rules-based CPQPredictika
A face velocity or sash opening the chosen series is not listed forOnly with care
Each series has its own lowest velocity and sash openings, so it takes a table or one rule per series, kept up to date by hand.
The series and the needs are in one request and compared with the specification on every call.
A special-purpose hood with the wrong material, or a liner the chemical attacksOnly with workarounds
Which liner suits which chemical is a 49-row table, and the perchloric and isotope hoods each need their own material rules; a rules engine needs an attribute per combination.
The materials and chemicals are part of the request and checked against the table on every call, with an approval for a marginal rating.
Duct collar or fire suppression counts that do not match the hood lengthOnly with care
Counts depend on the length, so they are separate rules that need updating when the specification changes.
The length and the stated counts are in the same request and compared on every call.
Every problem returned in one answerOnly with workarounds
Rules tend to fire as the order is saved or a step is completed, so a rep often fixes one error and meets the next.
All findings come back together, whatever order the choices arrived in.

What this validator catches: findings from its test cases

Click to expand

Every rule is checked on every call, in any order the choices arrive; one problem never hides another. Blocking findings stop the quote; approval findings need a person to sign off; everything else is computed.

1The application needs 50 FPM; the selected series is listed from 60 FPM upward.
2The application needs 60 FPM; the selected series is listed from 80 FPM upward.
3The application needs 80 FPM; the selected series is listed from 100 FPM upward.
4A face velocity of 70 FPM is not listed; the specification lists 50, 60, 80, 100 and 120 FPM.
5A sash opening of 35.0 inches is not listed for the selected series.
6A sash opening of 28.0 inches is not listed for the selected series.
7A hood length of 7 ft is not listed; the specification names hoods of 4, 5, 6, 8, 10 and 12 ft.
8Duct collars on the quote: 1; the selected hood length needs 2.
9The perchloric acid hood is specified with a Type 316L stainless steel liner.
10The perchloric acid hood is specified with a Type 316L stainless steel work surface.
11The perchloric acid hood is specified with a Type 316L stainless steel duct collar.
12The isotope hood is specified with a Type 304L stainless steel liner.
13The isotope hood is specified with a Type 304L stainless steel work surface.
14The isotope hood is specified with a Type 304L stainless steel duct collar.
15The standard hood duct collar is fiber reinforced plastic or Type 316 stainless steel.
16The specification rates the chemical to be used as D (swelling, pitting or severe etching) on the selected liner.
17The specification rates the chemical to be used as C (slight etching or severe staining) on the selected liner; a person should confirm the liner.
18Duct collars on the quote: 2; the selected hood length needs 1.
19Fire suppression units on the quote: 1; the selected hood length needs 2.
20The application needs 60 FPM; the selected series is listed from 100 FPM upward.

Scenarios and results

Click to expand

The recorded result of the validator for each scenario on this page: the request the AI agent sent, the verdict, and every finding, in the order they appear above.

1. Perchloric hood materials: A perchloric acid hood with a polyester liner, epoxy work surface and FRP duct collar

The agent's request. Please quote one Kewaunee Supreme Air Venturi hood, series V45, 6 ft long, at 100 FPM with a 28 inch sash opening, with a polyester liner, a epoxy work surface and a fiber reinforced plastic duct collar. The quote states 1 duct collar. It is ready to send.

Verdict: Blocked.

  • Perchloric hood needs a 316L liner. The perchloric acid hood is specified with a Type 316L stainless steel liner.
  • Perchloric hood needs a 316L work surface. The perchloric acid hood is specified with a Type 316L stainless steel work surface.
  • Perchloric hood needs a 316L duct collar. The perchloric acid hood is specified with a Type 316L stainless steel duct collar.

2. Perchloric hood materials: The agent resubmits after fixing the errors

The agent's request. Please quote one Kewaunee Supreme Air Venturi hood, series V45, 6 ft long, at 100 FPM with a 28 inch sash opening, with a 316L stainless steel liner, a 316L stainless steel work surface and a 316 stainless steel duct collar. The quote states 1 duct collar. It is ready to send.

Verdict: Allowed.

  • Hood summary. Selected hood, duct collars 1, listed from 100 FPM, sash opening 28.0 inches.

3. Hydrofluoric acid liner: A hood for hydrofluoric acid with a 304 stainless steel liner and two duct collars stated

The agent's request. Please quote one Kewaunee Supreme Air Venturi hood, series V06, 6 ft long, at 100 FPM with a 28 inch sash opening, with a 304 stainless steel liner, a 304 stainless steel work surface and a fiber reinforced plastic duct collar, to be used with 48 percent hydrofluoric acid. The quote states 2 duct collars. It is ready to send.

Verdict: Blocked.

  • Liner rated D for the chemical. The specification rates the chemical to be used as D (swelling, pitting or severe etching) on the selected liner.
  • Stated duct collars differ. Duct collars on the quote: 2; the selected hood length needs 1.

4. Hydrofluoric acid liner: The agent resubmits after fixing the errors

The agent's request. Please quote one Kewaunee Supreme Air Venturi hood, series V06, 6 ft long, at 100 FPM with a 28 inch sash opening, with a phenolic liner, a epoxy work surface and a fiber reinforced plastic duct collar, to be used with 48 percent hydrofluoric acid. The quote states 1 duct collar. It is ready to send.

Verdict: Allowed.

  • Hood summary. Selected hood, duct collars 1, listed from 50 FPM, sash opening 28.0 inches.

5. Chromic acid on a long LX hood: A 10 foot LX hood at 50 FPM for chromic acid with one fire suppression unit stated

The agent's request. Please quote one Kewaunee Supreme Air Venturi hood, series V26, 10 ft long, at 50 FPM with a 35 inch sash opening, with a 316L stainless steel liner, a 316L stainless steel work surface and a 316 stainless steel duct collar, to be used with 60 percent chromic acid, with fire suppression. The quote states 2 duct collars and 1 fire suppression unit. It is ready to send.

Verdict: Blocked.

  • Face velocity below the series range. The application needs 50 FPM; the selected series is listed from 60 FPM upward.
  • Stated fire suppression units differ. Fire suppression units on the quote: 1; the selected hood length needs 2.
  • Liner rated C for the chemical. The specification rates the chemical to be used as C (slight etching or severe staining) on the selected liner; a person should confirm the liner.

6. Chromic acid on a long LX hood: The agent resubmits after fixing the errors

The agent's request. Please quote one Kewaunee Supreme Air Venturi hood, series V26, 10 ft long, at 60 FPM with a 35 inch sash opening, with a 316L stainless steel liner, a 316L stainless steel work surface and a 316 stainless steel duct collar, to be used with 60 percent chromic acid, with fire suppression. The quote states 2 duct collars and 2 fire suppression units. It is ready to send.

Verdict: Needs approval.

  • Liner rated C for the chemical. The specification rates the chemical to be used as C (slight etching or severe staining) on the selected liner; a person should confirm the liner.
  • Hood summary. Selected hood, duct collars 2, listed from 60 FPM, sash opening 35.0 inches.

Narration transcript

Click to expand

The text of the narrated walkthrough, step by step.

  1. Welcome to a demo of Predictika's Validator. An AI sales agent prepares an order for a Kewaunee Supreme Air Venturi fume hood from Kewaunee's public guide specification. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model of the published specification and returns a verdict, every reason and the figures it computed.
  2. I will start with the first request by the agent.
  3. The agent says. Please quote one Kewaunee Supreme Air Venturi hood, series V45, 6 ft long, at 100 FPM with a 28 inch sash opening. The quote states 1 duct collar. The rest of the request is on screen.
  4. Before it goes out, the agent calls the validator.
  5. Blocked, with three findings, all in one answer. The perchloric acid hood is specified with a Type 316L stainless steel liner. The perchloric acid hood is specified with a Type 316L stainless steel work surface. The perchloric acid hood is specified with a Type 316L stainless steel duct collar.
  6. The validator returned three blocking findings in one answer. A check that reads one field at a time would pass most of this; it shows only when the choices in the quote are checked against each other.
  7. What this shows: for duct collar or fire suppression counts that do not match the hood length, a rules-based CPQ is error-prone for this.
  8. Now the agent resubmits.
  9. The corrected request. It changes the liner to 316L stainless steel, the work surface to 316L stainless steel and the duct collar to 316 stainless steel.
  10. It calls the validator again.
  11. This time the validator clears it.
  12. All three earlier findings are cleared.
  13. Three problems caught in one call, fixed in one pass, and a clean quote. That is the value of Predictika's validator: catching mistakes before they reach the customer. Now you can try the other scenarios yourself.