An AI sales agent prepares an order for a home or clinic ventilator: the device variant, its patient circuit, a humidifier, the options the customer asked for, the settings the patient needs and the battery hours it says the devices give. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model of the specifications and accessory notes Loewenstein Medical publishes and returns a verdict, every reason and the figures it computed.
This validator models an order for the prisma VENT50 or prisma VENT50-C ventilator, based on the public datasheets, accessories brochure and supply schedule, not just the scenarios shown below.
Any prisma VENT50 or VENT50-C order that touches these rules is in scope — not only the scenarios pictured here. Not covered: prices (the public supply schedule shows no price values, so this page shows no price build-up or line-item quote), other Loewenstein Medical devices such as the prisma VENT40, prisma VENT30-C, prisma30ST and LUISA, accessories not named here, and clinical decisions about the therapy. Assumptions: the humidifier voltage and modem region checks are generic matching logic, and the heated-circuit approval is read from the supply schedule's note that the 15 mm heated circuit is used with a temperature probe and heating wire.
7 scenarios in 3 groups · each group runs from the agent’s first attempt to the outcome · click one to switch
Illustrative agent statement. The validator receives the structured request below.
Proposing agent: AI sales agent (scripted)
Before sending, the agent must call Recorded run
What this means
Our assessment of what a typical rules-based CPQ (product rules, price rules and validation rules, as in Salesforce CPQ) does with the same cases. It is a judgment about effort and reliability, not a claim that nothing can be built.
| What the validator caught | Typical rules-based CPQ | Predictika |
|---|---|---|
| Settings limits that depend on the circuit: PEEP range with a leakage circuit against a single circuit with valve | Only with care The allowed PEEP range depends on which circuit is on another line, so a rule has to read the circuit and the setting together and be kept in step with the datasheet. | The circuit and every setting arrive in one request and are compared on every call. |
| Features that exist on only one variant: high flow and the heated leakage circuit on the VENT50-C | Only with workarounds The feature depends on the device variant on one line and a circuit or a flow setting on another, so it needs cross-line rules that interact with each other. | The variant, the circuit and the flow requirement are checked together, in any order. |
| Accessories that must be ordered separately: SpO2 sensor and cable, circuit holder | Only with care A separate-line requirement is easy to write for one module but has to be repeated for each, and a missing item is often caught only at the end or not at all. | Every item is in one request and each missing companion is reported with the rest. |
| Battery hours the order states | Only with care The stated figure is a free-entry field; a rule can compare it with a computed value, but each variant needs its own lookup. | The stated figure is a field in the request and is compared with the computed one on every call. |
| Every problem returned in one answer | Only with workarounds Rules tend to fire as the order is saved or a step is completed, so a person often fixes one error and meets the next. | All findings come back together, whatever order the choices arrived in. |
Every rule is checked on every call, in any order the choices arrive; one problem never hides another. Blocking findings stop the quote; approval findings need a person to sign off; everything else is computed.
| 1 | The customer needs an inspiratory pressure of 51 hPa; the device range is 4 to 50 hPa. |
| 2 | The customer needs an inspiratory pressure of 3 hPa; the device range is 4 to 50 hPa. |
| 3 | The customer needs PEEP of 0 hPa; with this circuit type the device range is 4 to 25 hPa. |
| 4 | The customer needs PEEP of 26 hPa; with this circuit type the device range is 0 to 25 hPa. |
| 5 | The customer needs a target volume of 99 ml; the device range is 100 to 2000 ml. |
| 6 | The customer needs a target volume of 2001 ml; the device range is 100 to 2000 ml. |
| 7 | The customer needs 16 l/min of oxygen feed; the maximum allowed flow with oxygen supply is 15 l/min. |
| 8 | High flow therapy is offered only on the prisma VENT50-C; this order has the prisma VENT50. |
| 9 | The customer needs high flow of 4 l/min; the prisma VENT50-C range is 5 to 60 l/min. |
| 10 | The customer needs high flow of 61 l/min; the prisma VENT50-C range is 5 to 60 l/min. |
| 11 | The heated leakage circuit is listed only for the prisma VENT50-C; this order has the prisma VENT50. |
| 12 | The heated leakage circuit is for high flow therapy only; the order has no high flow requirement. |
| 13 | Heated circuits are used with a temperature probe and heating wire, which come with the prisma VENT AQUA humidifier; a person should confirm the customer has one. |
| 14 | The 15 mm heated single circuit with valve can be used only for adults; this patient is a child. |
| 15 | A circuit diameter of 19 mm is not available; the circuits are 15 mm and 22 mm. |
| 16 | The humidifier voltage variant does not match the 115 V mains supply at the destination. |
| 17 | The prisma CLOUD modem variant does not match the destination region. |
| 18 | The prisma CHECK SpO2 module needs a sensor and a connection cable, which are ordered separately; the order is missing at least one. |
| 19 | The customer needs 8 hours of battery backup; this order has the device without the internal battery. |
| 20 | The customer needs 12 hours of backup; the datasheet gives more than 10 hours only at its stated test settings, so a person should confirm. |
The recorded result of the validator for each scenario on this page: the request the AI agent sent, the verdict, and every finding, in the order they appear above.
The agent's request. Please quote four prisma VENT50 ventilators with the internal battery, delivered to Europe, with a 22 mm leakage circuit. The patients need an inspiratory pressure of 52 hPa, PEEP of 3 hPa and 8 hours of battery backup. The quote states 12 hours of battery operation per device. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote four prisma VENT50 ventilators with the internal battery, delivered to Europe, with a 22 mm leakage circuit. The patients need an inspiratory pressure of 45 hPa, PEEP of 5 hPa and 8 hours of battery backup. The quote states 10 hours of battery operation per device. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote three prisma VENT50 ventilators with the internal battery, delivered to Europe, with a 22 mm heated leakage circuit, plus a 230 V prisma VENT AQUA humidifier. The patients need an inspiratory pressure of 25 hPa, PEEP of 6 hPa, high flow of 35 l/min and 6 hours of battery backup. The quote states 10 hours of battery operation per device. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote three prisma VENT50-C ventilators with the internal battery, delivered to Europe, with a 22 mm heated leakage circuit, plus a 230 V prisma VENT AQUA humidifier. The patients need an inspiratory pressure of 25 hPa, PEEP of 6 hPa, high flow of 35 l/min and 6 hours of battery backup. The quote states 10 hours of battery operation per device. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote three prisma VENT50 ventilators with the internal battery, delivered to North America, with a 22 mm single circuit with valve, plus the prisma CHECK SpO2 module and its connection cable, the 4G modem for prisma CLOUD in its EU variant and a Trolley 2.0 HC. The patients need an inspiratory pressure of 30 hPa, PEEP of 8 hPa and 8 hours of battery backup. The quote states 10 hours of battery operation per device. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote three prisma VENT50 ventilators with the internal battery, delivered to North America, with a 22 mm single circuit with valve, plus the prisma CHECK SpO2 module with a size M soft-tip sensor and its connection cable, the 4G modem for prisma CLOUD in its North America variant and a Trolley 2.0 HC. The patients need an inspiratory pressure of 30 hPa, PEEP of 8 hPa and 8 hours of battery backup. The quote states 10 hours of battery operation per device. It is ready to send.
Verdict: Needs approval.
The agent's request. Please quote three prisma VENT50 ventilators with the internal battery, delivered to North America, with a 22 mm single circuit with valve, plus the prisma CHECK SpO2 module with a size M soft-tip sensor and its connection cable, the 4G modem for prisma CLOUD in its North America variant and a Trolley 2.0 HC with a circuit holder. The patients need an inspiratory pressure of 30 hPa, PEEP of 8 hPa and 8 hours of battery backup. The quote states 10 hours of battery operation per device. It is ready to send.
Verdict: Allowed.
The text of the narrated walkthrough, step by step.