Predictika · Home ventilator orders

An AI sales agent's Loewenstein prisma VENT ventilator order, checked before it goes out

An AI sales agent prepares an order for a home-care provider: prisma VENT30-C, VENT40, VENT50 or VENT50-C ventilators, a ventilation mode, pressures, a patient circuit, humidification, the battery, modules such as prisma CHECK and prismaPSG, and the battery hours it states. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model of the public Loewenstein Medical documents plus ordering rules modeled from comparable home-ventilator programs, and returns a verdict, every reason and the figures it computed.

Scope: Loewenstein Medical prisma VENT home ventilators only (VENT30-C, VENT40, VENT50, VENT50-C): modes, pressure ranges, circuits, humidification, battery, modules and accessories. See everything this validator covers

This validator models orders for the Loewenstein Medical prisma VENT30-C, VENT40, VENT50 and VENT50-C ventilators, not just the scenarios shown below.

  • Model and pressures — the pressure range of each model (VENT30-C up to 30 hPa, VENT40 up to 40 hPa, VENT50 and VENT50-C up to 50 hPa), CPAP up to 20 hPa, PEEP ranges by circuit, respiratory rate, target volume of 100 to 2,000 ml and the 15 l/min oxygen feed limit.
  • Modes and circuits — autoS/T and mouthpiece ventilation as VENT50 and VENT50-C features, high flow (5 to 60 l/min) on the VENT50-C only, the single circuit with valve for VENT50 and VENT50-C, and the mouthpiece circuit at 15 mm.
  • Humidification and battery — the prismaAQUA humidifier and the heated circuits with auto-fill chamber (VENT50 and VENT50-C, adult use for the 15 mm valve circuit), and battery operation of up to 12 hours (VENT30-C, VENT40) or more than 10 hours (VENT50, VENT50-C) against the hours the order states.
  • Modules and accessories — prisma CHECK with the SpO2 sensor and cable ordered separately, prismaPSG with the communication module, and Trolley 2.0 with its separately ordered circuit arm. Reviews by a person apply to large orders (20 ventilators or more) and pediatric use (modeled).

Any prisma VENT30-C, VENT40, VENT50 or VENT50-C order that touches these rules is in scope — not only the scenarios pictured here. Not covered: prices (the price list is not modeled, so this page shows no price build-up), other Loewenstein ventilators and sleep therapy devices, the full accessory catalogue, and service terms. Several rules (PSG and communication module, high flow needing the heated circuit, pediatric and volume reviews, mode and circuit pairing) are modeled from comparable home-ventilator programs; they are not Loewenstein-published rules.

The AI sales agentBuilds the ventilator order from the conversation and says it is ready.
Predictika validatorChecks every rule at once and returns the verdict, each finding, and the computed figures.

7 scenarios in 3 groups · each group runs from the agent’s first attempt to the outcome · click one to switch

What the agent proposes

The AI sales agent says

Illustrative agent statement. The validator receives the structured request below.

Proposing agent: AI sales agent (scripted)

Before sending, the agent must call Recorded run

Predictika validator


      
Waiting for the call.

Where rules-based CPQ gets it wrong

Click to expand

Our assessment of what a typical rules-based CPQ (product rules, price rules and validation rules, as in Salesforce CPQ) does with the same cases. It is a judgment about effort and reliability, not a claim that nothing can be built.

What the validator caughtTypical rules-based CPQPredictika
Features that differ by model (modes, circuits, pressure ranges)Only with workarounds
Four models each have their own pressure range, modes and circuits. A rules-based CPQ needs a compatibility table, and every product update means editing it.
The model arrives in the request and its published ranges and features are checked on every call.
Mode, circuit and humidification togetherOnly with care
High flow, mouthpiece ventilation and the heated circuits each depend on the model and on each other. It can be built, but each new pairing is another rule to maintain.
Mode, circuit and humidification are checked together, in any order the choices arrive.
Stated battery hoursOnly with care
The figure depends on the model and on whether a battery is ordered, so a rule has to look it up and compare.
The stated hours are a field in the request, compared with the computed figure on every call.
Every problem returned in one answerOnly with workarounds
Rules tend to fire as the order is saved or a step is completed, so a person often fixes one error and meets the next.
All findings come back together, whatever order the choices arrived in.

What this validator catches: findings from its test cases

Click to expand

Every rule is checked on every call, in any order the choices arrive; one problem never hides another. Blocking findings stop the order; approval findings need a person to sign off; everything else is computed.

1A pressure of 35 hPa is outside the 4 to 30 hPa range of this ventilator model.
2A pressure of 45 hPa is outside the 4 to 40 hPa range of this ventilator model.
3A pressure of 3 hPa is outside the 4 to 50 hPa range of this ventilator model.
4CPAP pressure is limited to 20 hPa; the order has 25 hPa.
5With a leakage circuit PEEP is 4 to 25 hPa; the order has 3 hPa.
6With a single circuit with valve PEEP is 0 to 25 hPa; the order has 28 hPa.
7PEEP of 20 hPa must be below the inspiratory pressure of 20 hPa in this mode.
8A respiratory rate of 3 breaths per minute is outside the 5 to 60 range of this ventilator model.
9A respiratory rate of 70 breaths per minute is outside the 5 to 60 range of this ventilator model.
10A target volume of 50 ml is outside the 100 to 2,000 ml range.
11A target volume of 2,500 ml is outside the 100 to 2,000 ml range.
12The integrated oxygen port allows at most 15 l/min; the order has 20 l/min.
13AutoS/T mode is listed for prisma VENT50 and VENT50-C, not for this model.
14Mouthpiece ventilation is listed for prisma VENT50 and VENT50-C, not for this model.
15The mouthpiece circuit is a 15 mm circuit; the order has a 22 mm tube.
16Mouthpiece ventilation needs the mouthpiece circuit, and that circuit is used only in that mode.
17The high flow mode is offered only with prisma VENT50-C.
18High flow is 5 to 60 l/min; the order has 70 l/min.
19High flow therapy needs the heated circuit with auto-fill chamber.
20A high flow rate of 20 l/min is set but the mode is not high flow.
21The single circuit with valve is listed for prisma VENT50 and VENT50-C, not for this model.
22The heated circuits with auto-fill chamber are listed for prisma VENT50 and VENT50-C; this model uses the prismaAQUA humidifier.

Scenarios and results

Click to expand

The recorded result of the validator for each scenario on this page: the request the AI agent sent, the verdict, and every finding, in the order they appear above.

1. Home ventilators: Home ventilators for a care provider

The agent's request. Please order 8 prisma VENT30-C ventilators for adult use in autoS/T mode, on a single circuit with valve with a 22 mm tube, with no humidifier. Pressure is 35 hPa with PEEP 6 hPa, rate 15, target volume 500 ml. The order includes an internal battery. It states 12 hours of battery operation. It is ready to send.

Verdict: Blocked.

  • Pressure outside the model range. A pressure of 35 hPa is outside the 4 to 30 hPa range of this ventilator model.
  • autoS/T mode is a VENT50 / VENT50-C feature. AutoS/T mode is listed for prisma VENT50 and VENT50-C, not for this model.
  • Single circuit with valve is a VENT50 / VENT50-C circuit. The single circuit with valve is listed for prisma VENT50 and VENT50-C, not for this model.

2. Home ventilators: The agent resubmits after fixing the errors

The agent's request. Please order 8 prisma VENT50 ventilators for adult use in autoS/T mode, on a single circuit with valve with a 22 mm tube, with no humidifier. Pressure is 35 hPa with PEEP 6 hPa, rate 15, target volume 500 ml. The order includes an internal battery. It states 10 hours of battery operation. It is ready to send.

Verdict: Allowed.

  • Order summary. 8 ventilator(s); pressure range up to 50 hPa for this model; battery up to 10 hours.

3. High flow ventilators: High flow ventilators with monitoring

The agent's request. Please order 8 prisma VENT40 ventilators for adult use in high flow mode, on the high flow circuit with a 15 mm tube, with the prismaAQUA humidifier. Pressure is 20 hPa with PEEP 5 hPa, rate 15, target volume 500 ml, high flow 70 l/min. The order includes an internal battery, the prisma CHECK SpO2 module with 8 sensor(s) and 0 cable(s). It states 12 hours of battery operation. It is ready to send.

Verdict: Blocked.

  • High flow is a VENT50-C feature. The high flow mode is offered only with prisma VENT50-C.
  • High flow outside 5 to 60 l/min. High flow is 5 to 60 l/min; the order has 70 l/min.
  • High flow needs a heated circuit. High flow therapy needs the heated circuit with auto-fill chamber.
  • SpO2 module needs a sensor and a cable. The prisma CHECK SpO2 module needs a sensor and a connection cable ordered separately; the order has 8 sensor(s) and 0 cable(s).

4. High flow ventilators: The agent resubmits after fixing the errors

The agent's request. Please order 8 prisma VENT50-C ventilators for adult use in high flow mode, on the high flow circuit with a 15 mm tube, with the heated circuit with auto-fill chamber. Pressure is 20 hPa with PEEP 5 hPa, rate 15, target volume 500 ml, high flow 40 l/min. The order includes an internal battery, the prisma CHECK SpO2 module with 8 sensor(s) and 0 cable(s). It states 10 hours of battery operation. It is ready to send.

Verdict: Blocked.

  • SpO2 module needs a sensor and a cable. The prisma CHECK SpO2 module needs a sensor and a connection cable ordered separately; the order has 8 sensor(s) and 0 cable(s).

5. High flow ventilators: The agent resubmits after fixing the errors

The agent's request. Please order 8 prisma VENT50-C ventilators for adult use in high flow mode, on the high flow circuit with a 15 mm tube, with the heated circuit with auto-fill chamber. Pressure is 20 hPa with PEEP 5 hPa, rate 15, target volume 500 ml, high flow 40 l/min. The order includes an internal battery, the prisma CHECK SpO2 module with 8 sensor(s) and 8 cable(s). It states 10 hours of battery operation. It is ready to send.

Verdict: Allowed.

  • Order summary. 8 ventilator(s); pressure range up to 50 hPa for this model; battery up to 10 hours.

6. Pediatric ward rollout: A ventilator rollout for a pediatric ward

The agent's request. Please order 24 prisma VENT50 ventilators for pediatric use in ST (spontaneous/timed) mode, on a single circuit with valve with a 15 mm tube, with the heated circuit with auto-fill chamber. Pressure is 22 hPa with PEEP 6 hPa, rate 15, target volume 500 ml, oxygen feed 20 l/min. The order includes an internal battery, the prismaPSG module, Trolley 2.0. It states 10 hours of battery operation. It is ready to send.

Verdict: Blocked.

  • Oxygen feed above 15 l/min. The integrated oxygen port allows at most 15 l/min; the order has 20 l/min.
  • Heated 15 mm valve circuit is for adults only. The heated 15 mm single circuit with valve can be used only for adults.
  • PSG module needs the communication module. The prismaPSG module is connected through the prismaCONNECT communication module, which is not on the order.
  • Trolley 2.0 needs the circuit arm. The circuit arm for Trolley 2.0 has to be ordered separately and is not on the order.
  • Large ventilator order. An order of 24 ventilators needs a volume review by a person.
  • Pediatric use needs a clinical review. The order is for pediatric use; a person should confirm the settings and circuit with the clinical team.

7. Pediatric ward rollout: The agent resubmits after fixing the errors

The agent's request. Please order 24 prisma VENT50 ventilators for pediatric use in ST (spontaneous/timed) mode, on a single circuit with valve with a 22 mm tube, with the heated circuit with auto-fill chamber. Pressure is 22 hPa with PEEP 6 hPa, rate 15, target volume 500 ml, oxygen feed 10 l/min. The order includes an internal battery, the prismaPSG module, the prismaCONNECT communication module, Trolley 2.0 with the separate circuit arm. It states 10 hours of battery operation. It is ready to send.

Verdict: Needs approval.

  • Large ventilator order. An order of 24 ventilators needs a volume review by a person.
  • Pediatric use needs a clinical review. The order is for pediatric use; a person should confirm the settings and circuit with the clinical team.
  • Order summary. 24 ventilator(s); pressure range up to 50 hPa for this model; battery up to 10 hours.

Narration transcript

Click to expand

The text of the narrated walkthrough, step by step.

  1. Welcome to a demo of Predictika's Validator. An AI sales agent prepares an order for a home-care provider: prisma VENT30-C, VENT40, VENT50 or VENT50-C ventilators, and more. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model of the public Loewenstein Medical documents plus ordering rules modeled from comparable home-ventilator programs, and returns a verdict, every reason and the figures it computed.
  2. I will start with the first request by the agent.
  3. The agent says. Please order 8 prisma VENT30-C ventilators for adult use in autoS/T mode, on a single circuit with valve with a 22 mm tube, with no humidifier. Pressure is 35 hPa with PEEP 6 hPa, rate 15, target volume 500 ml. The rest of the request is on screen.
  4. Before it goes out, the agent calls the validator.
  5. Blocked, with three findings, all in one answer. A pressure of 35 hPa is outside the 4 to 30 hPa range of this ventilator model. AutoS/T mode is listed for prisma VENT50 and VENT50-C, not for this model. The single circuit with valve is listed for prisma VENT50 and VENT50-C, not for this model.
  6. The validator returned three blocking findings in one answer. A check that reads one field at a time would pass most of this; it shows only when the choices in the order are checked against each other.
  7. What this shows: for features that differ by model (modes, circuits, pressure ranges), a rules-based CPQ can do this only with workarounds.
  8. Now the agent resubmits.
  9. The corrected request. It changes the ventilator model to prisma VENT50 and the stated battery hours to 10.
  10. It calls the validator again.
  11. This time the validator clears it.
  12. All three earlier findings are cleared, and the validator's figure matches the order.
  13. Three problems caught in one call, fixed in one pass, and a clean order. That is the value of Predictika's validator: catching mistakes before they reach the customer. Now you can try the other scenarios yourself.