Predictika · Home ventilator orders
An AI sales agent prepares an order for a home-care provider: prisma VENT30-C, VENT40, VENT50 or VENT50-C ventilators, a ventilation mode, pressures, a patient circuit, humidification, the battery, modules such as prisma CHECK and prismaPSG, and the battery hours it states. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model of the public Loewenstein Medical documents plus ordering rules modeled from comparable home-ventilator programs, and returns a verdict, every reason and the figures it computed.
This validator models orders for the Loewenstein Medical prisma VENT30-C, VENT40, VENT50 and VENT50-C ventilators, not just the scenarios shown below.
Any prisma VENT30-C, VENT40, VENT50 or VENT50-C order that touches these rules is in scope — not only the scenarios pictured here. Not covered: prices (the price list is not modeled, so this page shows no price build-up), other Loewenstein ventilators and sleep therapy devices, the full accessory catalogue, and service terms. Several rules (PSG and communication module, high flow needing the heated circuit, pediatric and volume reviews, mode and circuit pairing) are modeled from comparable home-ventilator programs; they are not Loewenstein-published rules.
7 scenarios in 3 groups · each group runs from the agent’s first attempt to the outcome · click one to switch
Illustrative agent statement. The validator receives the structured request below.
Proposing agent: AI sales agent (scripted)
Before sending, the agent must call Recorded run
What this means
Our assessment of what a typical rules-based CPQ (product rules, price rules and validation rules, as in Salesforce CPQ) does with the same cases. It is a judgment about effort and reliability, not a claim that nothing can be built.
| What the validator caught | Typical rules-based CPQ | Predictika |
|---|---|---|
| Features that differ by model (modes, circuits, pressure ranges) | Only with workarounds Four models each have their own pressure range, modes and circuits. A rules-based CPQ needs a compatibility table, and every product update means editing it. | The model arrives in the request and its published ranges and features are checked on every call. |
| Mode, circuit and humidification together | Only with care High flow, mouthpiece ventilation and the heated circuits each depend on the model and on each other. It can be built, but each new pairing is another rule to maintain. | Mode, circuit and humidification are checked together, in any order the choices arrive. |
| Stated battery hours | Only with care The figure depends on the model and on whether a battery is ordered, so a rule has to look it up and compare. | The stated hours are a field in the request, compared with the computed figure on every call. |
| Every problem returned in one answer | Only with workarounds Rules tend to fire as the order is saved or a step is completed, so a person often fixes one error and meets the next. | All findings come back together, whatever order the choices arrived in. |
Every rule is checked on every call, in any order the choices arrive; one problem never hides another. Blocking findings stop the order; approval findings need a person to sign off; everything else is computed.
| 1 | A pressure of 35 hPa is outside the 4 to 30 hPa range of this ventilator model. |
| 2 | A pressure of 45 hPa is outside the 4 to 40 hPa range of this ventilator model. |
| 3 | A pressure of 3 hPa is outside the 4 to 50 hPa range of this ventilator model. |
| 4 | CPAP pressure is limited to 20 hPa; the order has 25 hPa. |
| 5 | With a leakage circuit PEEP is 4 to 25 hPa; the order has 3 hPa. |
| 6 | With a single circuit with valve PEEP is 0 to 25 hPa; the order has 28 hPa. |
| 7 | PEEP of 20 hPa must be below the inspiratory pressure of 20 hPa in this mode. |
| 8 | A respiratory rate of 3 breaths per minute is outside the 5 to 60 range of this ventilator model. |
| 9 | A respiratory rate of 70 breaths per minute is outside the 5 to 60 range of this ventilator model. |
| 10 | A target volume of 50 ml is outside the 100 to 2,000 ml range. |
| 11 | A target volume of 2,500 ml is outside the 100 to 2,000 ml range. |
| 12 | The integrated oxygen port allows at most 15 l/min; the order has 20 l/min. |
| 13 | AutoS/T mode is listed for prisma VENT50 and VENT50-C, not for this model. |
| 14 | Mouthpiece ventilation is listed for prisma VENT50 and VENT50-C, not for this model. |
| 15 | The mouthpiece circuit is a 15 mm circuit; the order has a 22 mm tube. |
| 16 | Mouthpiece ventilation needs the mouthpiece circuit, and that circuit is used only in that mode. |
| 17 | The high flow mode is offered only with prisma VENT50-C. |
| 18 | High flow is 5 to 60 l/min; the order has 70 l/min. |
| 19 | High flow therapy needs the heated circuit with auto-fill chamber. |
| 20 | A high flow rate of 20 l/min is set but the mode is not high flow. |
| 21 | The single circuit with valve is listed for prisma VENT50 and VENT50-C, not for this model. |
| 22 | The heated circuits with auto-fill chamber are listed for prisma VENT50 and VENT50-C; this model uses the prismaAQUA humidifier. |
The recorded result of the validator for each scenario on this page: the request the AI agent sent, the verdict, and every finding, in the order they appear above.
The agent's request. Please order 8 prisma VENT30-C ventilators for adult use in autoS/T mode, on a single circuit with valve with a 22 mm tube, with no humidifier. Pressure is 35 hPa with PEEP 6 hPa, rate 15, target volume 500 ml. The order includes an internal battery. It states 12 hours of battery operation. It is ready to send.
Verdict: Blocked.
The agent's request. Please order 8 prisma VENT50 ventilators for adult use in autoS/T mode, on a single circuit with valve with a 22 mm tube, with no humidifier. Pressure is 35 hPa with PEEP 6 hPa, rate 15, target volume 500 ml. The order includes an internal battery. It states 10 hours of battery operation. It is ready to send.
Verdict: Allowed.
The agent's request. Please order 8 prisma VENT40 ventilators for adult use in high flow mode, on the high flow circuit with a 15 mm tube, with the prismaAQUA humidifier. Pressure is 20 hPa with PEEP 5 hPa, rate 15, target volume 500 ml, high flow 70 l/min. The order includes an internal battery, the prisma CHECK SpO2 module with 8 sensor(s) and 0 cable(s). It states 12 hours of battery operation. It is ready to send.
Verdict: Blocked.
The agent's request. Please order 8 prisma VENT50-C ventilators for adult use in high flow mode, on the high flow circuit with a 15 mm tube, with the heated circuit with auto-fill chamber. Pressure is 20 hPa with PEEP 5 hPa, rate 15, target volume 500 ml, high flow 40 l/min. The order includes an internal battery, the prisma CHECK SpO2 module with 8 sensor(s) and 0 cable(s). It states 10 hours of battery operation. It is ready to send.
Verdict: Blocked.
The agent's request. Please order 8 prisma VENT50-C ventilators for adult use in high flow mode, on the high flow circuit with a 15 mm tube, with the heated circuit with auto-fill chamber. Pressure is 20 hPa with PEEP 5 hPa, rate 15, target volume 500 ml, high flow 40 l/min. The order includes an internal battery, the prisma CHECK SpO2 module with 8 sensor(s) and 8 cable(s). It states 10 hours of battery operation. It is ready to send.
Verdict: Allowed.
The agent's request. Please order 24 prisma VENT50 ventilators for pediatric use in ST (spontaneous/timed) mode, on a single circuit with valve with a 15 mm tube, with the heated circuit with auto-fill chamber. Pressure is 22 hPa with PEEP 6 hPa, rate 15, target volume 500 ml, oxygen feed 20 l/min. The order includes an internal battery, the prismaPSG module, Trolley 2.0. It states 10 hours of battery operation. It is ready to send.
Verdict: Blocked.
The agent's request. Please order 24 prisma VENT50 ventilators for pediatric use in ST (spontaneous/timed) mode, on a single circuit with valve with a 22 mm tube, with the heated circuit with auto-fill chamber. Pressure is 22 hPa with PEEP 6 hPa, rate 15, target volume 500 ml, oxygen feed 10 l/min. The order includes an internal battery, the prismaPSG module, the prismaCONNECT communication module, Trolley 2.0 with the separate circuit arm. It states 10 hours of battery operation. It is ready to send.
Verdict: Needs approval.
The text of the narrated walkthrough, step by step.