Predictika · Patient monitor configuration: bedside monitors and parameter modules
An AI sales agent prepares a configuration for a bedside patient monitor: the monitor, an optional extension and the parameter modules. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model of the public product information and returns a verdict, every reason and the figures it computed.
This validator models configurations of one product family, the BeneVision N12, N15 and N17 bedside monitors with N-Series parameter modules, not just the scenarios shown below.
Any N12, N15 or N17 configuration that touches these rules is in scope — not only the scenarios pictured here. Not covered: prices (Mindray does not publish them, so this validator produces no price build-up or line-item quote), the N1, N19 and N22 monitors, satellite module racks, the recorder, software options, central station and other Mindray product lines.
7 scenarios in 3 groups · each group runs from the agent’s first attempt to the outcome · click one to switch
Illustrative agent statement. The validator receives the structured request below.
Proposing agent: AI sales agent (scripted)
Before sending, the agent must call Recorded run
What this means
Our assessment of what a typical rules-based CPQ (product rules, price rules and validation rules, as in Salesforce CPQ) does with the same cases. It is a judgment about effort and reliability, not a claim that nothing can be built.
| What the validator caught | Typical rules-based CPQ | Predictika |
|---|---|---|
| Modules that take two slots and others that take one | Only with care A product rule per module with its slot width and a running total against the monitor; it works, but the total and the message must be kept in step as modules are added. | Module counts arrive in one request and the slots needed are computed and compared with the slots available on every call. |
| An extension that changes how many slots a monitor has | Only with care The extension choice changes the slot limit differently for each monitor, so each monitor and extension combination needs its own rule. | Monitor, extension and modules are in one request and the available slots are computed from the monitor and extension together. |
| Modules that are not for every patient population | Only with workarounds A module rule has to read the patient population chosen elsewhere in the configuration, and each restricted module needs its own rule and message. | The patient population travels with the configuration and each restricted module is checked against it. |
| Stated slot count that must match the modules | Only with care A stated figure is a free-entry field; a rule can compare it with a computed total, but someone has to write and maintain that comparison. | The stated figure travels with the configuration and is compared with the computed figure on every call. |
| Every problem returned in one answer | Only with workarounds Rules tend to fire as the order is saved or a step is completed, so a rep often fixes one error and meets the next. | All findings come back together, whatever order the choices arrived in. |
Every rule is checked on every call, in any order the choices arrive; one problem never hides another. Blocking findings stop the quote; approval findings need a person to sign off; everything else is computed.
| 1 | The modules need 10 slots; the monitor with the chosen extension has 9. |
| 2 | The modules need 8 slots; the monitor with the chosen extension has 6. |
| 3 | The modules need 5 slots; the monitor with the chosen extension has 4. |
| 4 | The modules need 8 slots; the monitor with the chosen extension has 7. |
| 5 | Mindray lists only the TMR extension (7 slots in total) for the N12; the SMR is listed for the N15 and N17. |
| 6 | The BIS module is for adult and pediatric patients only; the quote is for neonatal use. |
| 7 | The CO module is for adult patients only. |
| 8 | The FloTrac module is for adult patients only. |
| 9 | The quote states 7 slots in use; the modules need 9. |
| 10 | The quote states 10 slots in use; the modules need 9. |
The recorded result of the validator for each scenario on this page: the request the AI agent sent, the verdict, and every finding, in the order they appear above.
The agent's request. Please configure a BeneVision N12 bedside monitor with the SMR extension for adult patients, with 1 Platinum MPM, 1 Agent Gas module, 1 CO module, 1 FloTrac module and 3 other single-slot modules. The configuration states 9 module slots in use. It is ready to send.
Verdict: Blocked.
The agent's request. Please configure a BeneVision N12 bedside monitor with the TMR extension for adult patients, with 1 Platinum MPM, 1 Agent Gas module, 1 CO module, 1 FloTrac module and 3 other single-slot modules. The configuration states 9 module slots in use. It is ready to send.
Verdict: Blocked.
The agent's request. Please configure a BeneVision N12 bedside monitor with the TMR extension for adult patients, with 1 Platinum MPM, 1 Agent Gas module, 1 CO module and 2 other single-slot modules. The configuration states 7 module slots in use. It is ready to send.
Verdict: Allowed.
The agent's request. Please configure a BeneVision N15 bedside monitor without an extension for adult patients, with 1 Platinum MPM, 1 Agent Gas module, 1 BIS module, 1 CO module, 1 FloTrac module and 2 other single-slot modules. The configuration states 9 module slots in use. It is ready to send.
Verdict: Blocked.
The agent's request. Please configure a BeneVision N15 bedside monitor with the TMR extension for adult patients, with 1 Platinum MPM, 1 Agent Gas module, 1 BIS module, 1 CO module, 1 FloTrac module and 2 other single-slot modules. The configuration states 9 module slots in use. It is ready to send.
Verdict: Allowed.
The agent's request. Please configure a BeneVision N12 bedside monitor with the TMR extension for neonatal patients, with 1 Platinum MPM, 1 BIS module, 1 CO module and 2 other single-slot modules. The configuration states 6 module slots in use. It is ready to send.
Verdict: Blocked.
The agent's request. Please configure a BeneVision N12 bedside monitor with the TMR extension for neonatal patients, with 1 Platinum MPM and 4 other single-slot modules. The configuration states 6 module slots in use. It is ready to send.
Verdict: Allowed.
The text of the narrated walkthrough, step by step.