Predictika · Neurodiagnostic EEG amplifier orders
An AI sales agent prepares an order for an EEG amplifier and NeuroWorks software from public product pages, with the channels, DC inputs, sampling rate and software version a study needs. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model of the published specifications and returns a verdict, every reason and the figures it computed.
This validator models orders for three Natus EEG amplifiers used with NeuroWorks software, based on the public Natus product pages, not just the scenarios shown below.
Any order for these amplifiers that touches these rules is in scope — not only the scenarios pictured here. Not covered: prices (none are published, so this validator produces no price build-up or line-item quote), the rest of the Natus range (sleep, newborn care, EMG), accessories, service and licences.
7 scenarios in 3 groups · each group runs from the agent’s first attempt to the outcome · click one to switch
Illustrative agent statement. The validator receives the structured request below.
Proposing agent: AI sales agent (scripted)
Before sending, the agent must call Recorded run
What this means
Our assessment of what a typical rules-based CPQ (product rules, price rules and validation rules, as in Salesforce CPQ) does with the same cases. It is a judgment about effort and reliability, not a claim that nothing can be built.
| What the validator caught | Typical rules-based CPQ | Predictika |
|---|---|---|
| An amplifier that does not support the study's channel count or sampling rate | Only with care Each amplifier has its own channel, DC and sampling limits, so each limit becomes a separate rule per model that must be kept in step with the product pages. | The study requirements and the amplifier are in one request and compared with every limit on every call. |
| An amplifier listed for one application but not another | Only with care Which amplifier suits routine EEG, long-term monitoring or ICU is a compatibility table, usually maintained by hand per product. | The amplifier and the application are checked together against the published list on every call. |
| A software option that needs a newer NeuroWorks version | Only with workarounds The option (autoSCORE) and the amplifier each depend on a software version, so the rules span several products and versions and break quietly when one changes. | The version is part of the request and is compared with each requirement on every call. |
| A stated capacity that disagrees with the product page | Only with workarounds The figure the agent states is free text or a separate field, so it takes an extra rule that compares it with the amplifier's own capacity. | The stated figure is part of the request and is compared with the published capacity on every call. |
Every rule is checked on every call, in any order the choices arrive; one problem never hides another. Blocking findings stop the quote; approval findings need a person to sign off; everything else is computed.
| 1 | The study needs 257 EEG channels; the selected amplifier offers at most 256 (Quantum 256, EMU40EX 40, Brain Monitor 64). |
| 2 | The study needs a sampling rate of 16,001 Hz; the selected amplifier samples at up to 16,000 Hz (Quantum 16 kHz, EMU40EX 1,024 Hz, Brain Monitor 4 kHz). |
| 3 | The study needs 17 DC inputs; the selected amplifier offers at most 16 (Quantum 16, EMU40EX 4, Brain Monitor 16). |
| 4 | The quote has NeuroWorks version 8; the selected amplifier requires NeuroWorks 9 or newer. |
| 5 | The NeuroWorks page does not list the selected amplifier for the selected application (routine EEG: Brain Monitor; long-term monitoring: Quantum, EMU40EX, Brain Monitor; ICU: Brain Monitor, EMU40EX). |
| 6 | The study needs 41 EEG channels; the selected amplifier offers at most 40 (Quantum 256, EMU40EX 40, Brain Monitor 64). |
| 7 | The study needs 5 DC inputs; the selected amplifier offers at most 4 (Quantum 16, EMU40EX 4, Brain Monitor 16). |
| 8 | The study needs a sampling rate of 2,048 Hz; the selected amplifier samples at up to 1,024 Hz (Quantum 16 kHz, EMU40EX 1,024 Hz, Brain Monitor 4 kHz). |
The recorded result of the validator for each scenario on this page: the request the AI agent sent, the verdict, and every finding, in the order they appear above.
The agent's request. Please quote a Natus EMU40EX wireless LTM amplifier for long-term monitoring, with 48 EEG channels and 2 DC inputs sampled at 512 Hz, on NeuroWorks version 9. The quote states a maximum of 48 channels for this amplifier. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote a Natus Quantum LTM amplifier for long-term monitoring, with 48 EEG channels and 2 DC inputs sampled at 512 Hz, on NeuroWorks version 9. The quote states a maximum of 256 channels for this amplifier. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote a Natus Brain Monitor EEG amplifier for routine EEG, with 40 EEG channels and 4 DC inputs sampled at 1,000 Hz, on NeuroWorks version 8 with autoSCORE. The quote states a maximum of 64 channels for this amplifier. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote a Natus Brain Monitor EEG amplifier for routine EEG, with 40 EEG channels and 4 DC inputs sampled at 1,000 Hz, on NeuroWorks version 9 with autoSCORE. The quote states a maximum of 64 channels for this amplifier. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote a Natus Brain Monitor EEG amplifier for routine EEG, with 40 EEG channels and 4 DC inputs sampled at 1,000 Hz, on NeuroWorks version 10 with autoSCORE. The quote states a maximum of 64 channels for this amplifier. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote a Natus Quantum LTM amplifier for ICU monitoring, with 48 EEG channels and 4 DC inputs sampled at 2,000 Hz, on NeuroWorks version 9. The quote states a maximum of 256 channels for this amplifier. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote a Natus Brain Monitor EEG amplifier for ICU monitoring, with 48 EEG channels and 4 DC inputs sampled at 2,000 Hz, on NeuroWorks version 9. The quote states a maximum of 64 channels for this amplifier. It is ready to send.
Verdict: Allowed.