Predictika · Apheresis procedure and disposable orders
An AI sales agent prepares an order for a hospital apheresis service that runs the Terumo BCT Spectra Optia Apheresis System: the procedure, the matching disposable set, the patient entries, the anticoagulant ratio, any custom prime, the access type and the circuit volume it states. Before it goes out, the agent sends it to Predictika, which checks it against a declarative model built from public Terumo BCT information plus rules modeled from comparable apheresis workflows, and returns a verdict, every reason and the figures it computed.
This validator models Spectra Optia procedure and disposable-set orders, not just the scenarios shown below. It is a demonstration of order checking, not clinical advice.
Any Spectra Optia order that touches these rules is in scope — not only the scenarios pictured here. Not covered: prices (Terumo BCT does not publish list prices, so this page shows no price build-up), Trima Accel and other Terumo BCT systems, other set variants such as Correct Connect or low-volume sets, software versions, and any clinical judgment. Public from Terumo BCT: the procedure names, which set is used for TPE, RBC exchange and CMNC, the minimum patient entries, the under-25 kg blood volume note, the default 12 to 1 ratio and the albumin or RBC prime. Modeled by this page, not Terumo-published: the set circuit volumes (190 and 170 mL), the 15 percent circuit volume rule, the 8 to 25 ratio range and the approval above 15, the 20 percent hematocrit floor, single-needle use, the approval for patients under 25 kg and for RBC primes.
7 scenarios in 3 groups · each group runs from the agent’s first attempt to the outcome · click one to switch
Illustrative agent statement. The validator receives the structured request below.
Proposing agent: AI sales agent (scripted)
Before sending, the agent must call Recorded run
What this means
Our assessment of what a typical rules-based CPQ (product rules, price rules and validation rules, as in Salesforce CPQ) does with the same cases. It is a judgment about effort and reliability, not a claim that nothing can be built.
| What the validator caught | Typical rules-based CPQ | Predictika |
|---|---|---|
| Procedure, disposable set and settings that must agree | Only with care Each procedure uses a particular set, and prime, access and hematocrit options differ by procedure. It can be built, but each new procedure or set variant adds rules to maintain. | The procedure, set, prime, access and settings are in one request and are checked together. |
| Circuit volume against the patient size | Only with care The volume depends on the chosen set and is compared with a share of the entered blood volume, so a rule has to recompute it at every change. | The set volume and the 15 percent threshold are computed on every call, including at the boundary. |
| Stated circuit volume | Only with care The stated volume is derived from the chosen set, so a rule has to look it up and compare. | The stated volume is a field in the request, compared with the set's volume on every call. |
| Every problem returned in one answer | Only with workarounds Rules tend to fire as the order is saved or a step is completed, so a person often fixes one error and meets the next. | All findings come back together, whatever order the choices arrived in. |
Every rule is checked on every call, in any order the choices arrive; one problem never hides another. Blocking findings stop the quote; approval findings need a person to sign off; everything else is computed.
| 1 | The selected disposable set is not the set used for this procedure; plasma exchange, RBC exchange and RBC depletion use the Exchange Set, and MNC and CMNC collections use the IDL Set. |
| 2 | A patient weight of 1 kg is below the 2 kg minimum the system accepts. |
| 3 | A patient of 1 kg is in the low-blood-volume range; a clinician should confirm the order. |
| 4 | A patient height of 29 cm is below the 30 cm minimum the system accepts. |
| 5 | A total blood volume of 299 mL is below the 300 mL minimum the system accepts. |
| 6 | The system does not calculate total blood volume for patients under 25 kg; it must be entered manually. |
| 7 | A patient of 24 kg is in the low-blood-volume range; a clinician should confirm the order. |
| 8 | An inlet to anticoagulant ratio of 7 to 1 is outside the 8 to 25 range this model allows. |
| 9 | An inlet to anticoagulant ratio of 16 to 1 reduces circuit anticoagulation and raises the risk of clumping; a person should confirm it. |
| 10 | An inlet to anticoagulant ratio of 25 to 1 reduces circuit anticoagulation and raises the risk of clumping; a person should confirm it. |
| 11 | An inlet to anticoagulant ratio of 26 to 1 is outside the 8 to 25 range this model allows. |
| 12 | An RBC prime is offered only for RBC exchange and RBC depletion. |
| 13 | An RBC prime needs a compatible RBC unit confirmed by the blood bank. |
| 14 | A custom prime is modeled for the Exchange Set only; the IDL Set is run without one. |
| 15 | The extracorporeal volume of 190 mL is more than 15 percent of the blood volume; a custom prime with albumin or RBC is needed. |
| 16 | Single-needle access is modeled for exchange and depletion procedures, not for MNC or CMNC collections. |
| 17 | A target hematocrit of 19 percent is below the 20 percent floor this model allows. |
| 18 | An RBC depletion cannot raise the hematocrit; the target is 30 percent and the patient is at 25 percent. |
| 19 | The order states an extracorporeal volume of 175 mL; the selected set gives 190 mL. |
The recorded result of the validator for each scenario on this page: the request the AI agent sent, the verdict, and every finding, in the order they appear above.
The agent's request. Please quote the Spectra Optia CMNC collection with the IDL Set for a patient of 26 kg and 125 cm with 1,800 mL of blood volume, using an inlet to anticoagulant ratio of 30 to 1, no custom prime and single-needle access. The order states an extracorporeal volume of 170 mL. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote the Spectra Optia CMNC collection with the IDL Set for a patient of 26 kg and 125 cm with 1,800 mL of blood volume, using an inlet to anticoagulant ratio of 20 to 1, no custom prime and standard access. The order states an extracorporeal volume of 170 mL. It is ready to send.
Verdict: Needs approval.
The agent's request. Please quote the Spectra Optia CMNC collection with the IDL Set for a patient of 26 kg and 125 cm with 1,800 mL of blood volume, using an inlet to anticoagulant ratio of 12 to 1, no custom prime and standard access. The order states an extracorporeal volume of 170 mL. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote the Spectra Optia RBC exchange with the IDL Set for a patient of 30 kg and 130 cm with 2,100 mL of blood volume to a target hematocrit of 15 percent (the patient is at 25 percent), using an inlet to anticoagulant ratio of 12 to 1, no custom prime and standard access. The order states an extracorporeal volume of 170 mL. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote the Spectra Optia RBC exchange with the Exchange Set for a patient of 30 kg and 130 cm with 2,100 mL of blood volume to a target hematocrit of 30 percent (the patient is at 25 percent), using an inlet to anticoagulant ratio of 12 to 1, no custom prime and standard access. The order states an extracorporeal volume of 190 mL. It is ready to send.
Verdict: Allowed.
The agent's request. Please quote the Spectra Optia therapeutic plasma exchange (TPE) with the Exchange Set for a patient of 25 kg and 120 cm with 1,200 mL of blood volume, using an inlet to anticoagulant ratio of 12 to 1, no custom prime and standard access. The order states an extracorporeal volume of 175 mL. It is ready to send.
Verdict: Blocked.
The agent's request. Please quote the Spectra Optia therapeutic plasma exchange (TPE) with the Exchange Set for a patient of 25 kg and 120 cm with 1,200 mL of blood volume, using an inlet to anticoagulant ratio of 12 to 1, an albumin prime and standard access. The order states an extracorporeal volume of 190 mL. It is ready to send.
Verdict: Allowed.
The text of the narrated walkthrough, step by step.